Metallosis following implantation of magnetically controlled growing rods in the treatment of scoliosis
- 1 December 2016
- journal article
- case report
- Published by British Editorial Society of Bone & Joint Surgery in The Bone & Joint Journal
- Vol. 98-B (12) , 1662-1667
Abstract
Aims: We present a case series of five patients who had revision surgery following magnetic controlled growing rods (MGCR) for early onset scoliosis. Metallosis was found during revision in four out of five patients and we postulated a mechanism for rod failure based on retrieval analysis. Patients and Methods: Retrieval analysis was performed on the seven explanted rods. The mean duration of MCGR from implantation to revision was 35 months (17 to 46). The mean age at revision was 12 years (7 to 15; four boys, one girl). Results: A total of six out of seven rods had tissue metallosis and pseudo-capsule surrounding the actuator. A total of four out of seven rods were pistoning. There were two rods which were broken. All rods had abrasive circumferential markings. A significant amount of metal debris was found when the actuators were carefully cut open. Analytical electron microscopy demonstrated metal fragments of predominantly titanium with a mean particle size of 3.36 microns (1.31 to 6.61). Conclusion: This study highlights concerns with tissue metallosis in MCGR. We recommend careful follow-up of patients who have received this implant. Cite this article: Bone Joint J 2016;98-B:1662–7.Keywords
This publication has 10 references indexed in Scilit:
- Early Onset Scoliosis Consensus Statement, SRS Growing Spine Committee, 2015Spine Deformity, 2015
- Early experience of MAGEC magnetic growing rods in the treatment of early onset scoliosisEuropean Spine Journal, 2014
- Next Generation of Growth-Sparing TechniquesSpine, 2013
- Early results of a remotely-operated magnetic growth rod in early-onset scoliosisThe Bone & Joint Journal, 2013
- Magnetically controlled growing rods for severe spinal curvature in young children: a prospective case seriesThe Lancet, 2012
- Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and WalesThe Lancet, 2012
- Pseudotumors Associated with Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2012
- Complications of Growing-Rod Treatment for Early-Onset ScoliosisJournal of Bone and Joint Surgery, 2010
- Comparison of Complications Among Growing Spinal ImplantsSpine, 2010
- General anesthesia, surgery and hospitalization in children and their effects upon cognitive, academic, emotional and sociobehavioral development – a reviewPediatric Anesthesia, 2004