Finite Element Study to Evaluate the Biomechanical Performance of the Spine After Augmenting Percutaneous Pedicle Screw Fixation With Kyphoplasty in the Treatment of Burst Fractures
- 21 March 2018
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 140 (6)
Abstract
Percutaneous pedicle screw fixation (PPSF) is a well-known minimally invasive surgery (MIS) employed in the treatment of thoracolumbar burst fractures (TBF). However, hardware failure and loss of angular correction are common limitations caused by the poor support of the anterior column of the spine. Balloon kyphoplasty (KP) is another MIS that was successfully used in the treatment of compression fractures by augmenting the injured vertebral body with cement. To overcome the limitations of stand-alone PPSF, it was suggested to augment PPSF with KP as a surgical treatment of TBF. Yet, little is known about the biomechanical alteration occurred to the spine after performing such procedure. The objective of this study was to evaluate and compare the immediate post-operative biomechanical performance of stand-alone PPSF, stand-alone-KP, and KP-augmented PPSF procedures. Novel three-dimensional (3D) finite element (FE) models of the thoracolumbar junction that describes the fractured spine and the three investigated procedures were developed and tested under mechanical loading conditions. The spinal stiffness, stresses at the implanted hardware, and the intradiscal pressure at the upper and lower segments were measured and compared. The results showed no major differences in the measured parameters between stand-alone PPSF and KP-augmented PPSF procedures, and demonstrated that the stand-alone KP may restore the stiffness of the intact spine. Accordingly, there was no immediate post-operative biomechanical advantage in augmenting PPSF with KP when compared to stand-alone PPSF, and fatigue testing may be required to evaluate the long-term biomechanical performance of such procedures.Keywords
This publication has 48 references indexed in Scilit:
- FEBio: Finite Elements for BiomechanicsJournal of Biomechanical Engineering, 2012
- Percutaneous kyphoplasty and pedicle screw fixation for the management of thoraco-lumbar burst fracturesEuropean Spine Journal, 2010
- Effect of multilevel lumbar disc arthroplasty on spine kinematics and facet joint loads in flexion and extension: a finite element analysisEuropean Spine Journal, 2010
- Kyphoplasty for the treatment of incomplete osteoporotic burst fracturesEuropean Spine Journal, 2010
- Short segment pedicle screw instrumentation and augmentation vertebroplasty in lumbar burst fractures: an experienceEuropean Spine Journal, 2008
- Biomechanical Rationale for Using Polyetheretherketone (PEEK) Spacers for Lumbar Interbody Fusion–A Finite Element StudySpine, 2006
- Prospective study of standalone balloon kyphoplasty with calcium phosphate cement augmentation in traumatic fracturesEuropean Spine Journal, 2006
- Application of a new calibration method for a three-dimensional finite element model of a human lumbar annulus fibrosusClinical Biomechanics, 2006
- Biomechanical properties of human lumbar spine ligamentsJournal of Biomechanics, 1992
- Mechanical Response of a Lumbar Motion Segment in Axial Torque Alone and Combined with CompressionSpine, 1986