The error due to fat inhomogeneity in lumbar spine bone mineral measurements
Authors:
- 1 February 1989
- journal article
- Published by IOP Publishing in Clinical Physics and Physiological Measurement
- Vol. 10 (1) , 57-64
Abstract
Bone mineral mass measurements using dual photon absorptiometry rely on the assumption that fat thickness is constant within the region scanned by the photon beam. This assumption has been tested using both single and multiple slices from archived CT scans. In 26 patients, the difference in fat content between the bone and baseline measurement sites, the fat deviation, was found to vary from -2.7 to 18.7%. The median fat deviation was 4.4% which can lead to errors in measured lumbar spine bone mineral of 6 to 11%. Approximately one third of patients have errors greater than 10%.Keywords
This publication has 13 references indexed in Scilit:
- Methodologies for the measurement of bone density and their precision and accuracySeminars in Nuclear Medicine, 1987
- The effect of fat on bone mineral measurements in normal subjects with recommended values of bone, muscle and fat attenuation coefficientsClinical Physics and Physiological Measurement, 1987
- Photon absorptiometric studies of elements, mixtures and substances of biomedical interestPhysics in Medicine & Biology, 1986
- Dual-photon Gd-153 absorptiometry of bone.Radiology, 1985
- Photon mass attenuation and energy-absorption coefficientsThe International Journal of Applied Radiation and Isotopes, 1982
- Postmenopausal osteoporosis.BMJ, 1982
- Measurement of bone mineral content in human vertebrae and hip by dual photon absorptiometry.Radiology, 1980
- Dual Photon Absorptiometry in Lumbar Vertebrae Evaluation of the baseline errorActa Radiologica: Oncology, 1980
- ATTENUATION OF MONOENERGETIC GAMMA RAYS IN TISSUESAmerican Journal of Roentgenology, 1975
- Determination of the Electronic density and the average atomic number of tissues in man by γ-ray attenuationCellular and Molecular Life Sciences, 1974