Phrenic Stimulation

Authors:

Abstract
Background— Phrenic stimulation (PS) may hinder left ventricular (LV) pacing. We prospectively observed its prevalence in consecutive patients with cardiac resynchronization therapy (CRT) devices. Methods and Results— In the years 2003 to 2006, 197 patients received a CRT device. PS and LV threshold measurements were carried out at implantation and at 6-month follow-up. LV reverse remodeling was assessed by echocardiography before implantation and at follow-up. LV lead placement was lateral/posterolateral in 86% of patients. Both PS and LV reverse remodeling occurred most frequently at the lateral/posterolateral LV pacing sites ( P Conclusions— PS may seriously hinder CRT. A bipolar LV lead and cathode programmability are mandatory to avoid PS by changing the LV pacing vector at target sites for CRT. LV stability at target sites despite PS should also be pursued by these means. The automatic adjustment of LV pacing output is complementary in patients with a small PS-LV threshold difference.