Validity and reliability of Gripwise digital dynamometer in the assessment of handgrip strength in older adults
Open Access
- 25 April 2025
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Aging
- Vol. 6, 1560097
Abstract
Introduction: With the advancement of studies on the importance of sarcopenia in the aging process, new technologies have been developed to assess muscle mass and function. However, most research on portable devices has not considered a wide range of ages and clinical conditions. This study aimed to evaluate the reliability of the Gripwise digital dynamometer in measuring handgrip strength in older Brazilian adults, comparing its performance with the widely used Saehan device.Methods: A cross-sectional study was conducted with 149 participants (32 men and 117 women), with an average age of 69.5 years. Handgrip strength was measured using both the Gripwise and Saehan dynamometers. Reliability was assessed using the intraclass correlation coefficient (ICC). Analyses considered three handgrip strength measurements from both devices, as well as the highest value obtained. The classification of dynapenia (low muscle strength) was compared using different cutoff points proposed by Villain et al. (2023), Spexoto et al. (2022), and Cruz-Jentoft et al. (2019).Results: Both dynamometers demonstrated excellent reliability, with ICC values above 0.90. However, significant differences in mean handgrip strength values were observed between the devices (approximately 3.5–four kgf). These variations impacted the classification of dynapenia, with the Gripwise identifying more cases of low muscle strength compared to Saehan.Conclusion: The lower values reported by the Gripwise may impact clinical decision-making in two ways. On one hand, lower values may lead to earlier detection of muscle weakness, allowing for quicker intervention in individuals with strength below typical thresholds. However, this could also result in an overestimation of the prevalence of dynapenia if the values do not accurately reflect true muscle strength, which could lead to unnecessary interventions. Therefore, it is crucial to consider the need for adjustments in the cutoff points when using Gripwise. These findings highlight the need to revise cutoff points for dynapenia classification, considering device variations and model differences in older age groups.Keywords
This publication has 28 references indexed in Scilit:
- International norms for adult handgrip strength: A systematic review of data on 2.4 million adults aged 20 to 100+ years from 69 countries and regionsJournal of Sport and Health Science, 2024
- The use of portable A-mode ultrasound in appendicular lean mass measurements among older adults: a comparison study with dual-energy X-ray absorptiometry and handgrip strengthEuropean Journal of Clinical Nutrition, 2024
- Sarcopenia and Sarcopenic Obesity and Mortality Among Older PeopleJAMA Network Open, 2024
- Intra- and inter-rater reliability of muscle and fat thickness measurements obtained using portable ultrasonography in older adultsClinical Nutrition ESPEN, 2024
- Handgrip strength is positively associated with successful aging in older adults: A national cross-sectional study in ChinaJournal of Affective Disorders, 2023
- Handgrip Measurement Method Affects Asymmetry but Not Weakness Identification in Community-Dwelling Older AdultsJournal of the American Medical Directors Association, 2022
- A Novel Dynamometer: The Gripwise2022
- Quantification of Scan Analysis Errors in GE Lunar DXA Visceral Adiposity in AdultsJournal of Clinical Densitometry, 2021
- The Short-Term Reliability of Grip Strength Measurement and the Effects of Posture and Grip SpanThe Journal of Hand Surgery, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986