Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions
Open Access
- 5 September 2013
- journal article
- research article
- Published by Springer Nature in BMC Ophthalmology
- Vol. 13 (1) , 44
Abstract
Background: Icare PRO (ICP) is a new Rebound tonometer that is able to measure intraocular pressure (IOP) in both sitting and reclining positions. In this study, the gold standard Goldmann tonometer (GAT) was compared to ICP and Tono-Pen AVIA (TPA). Hypothesis was that repeatability of GAT is superior to ICP and TPA. Methods: 36 eyes of 36 healthy caucasian individuals, 13 male and 26 females, 17 right and 19 left eyes have been included in this prospective, randomized, cross-sectional study. The study was conducted at a single site (Dept. of Ophthalmology, UniversityHospital Zurich, Switzerland). Primary outcome measures were Intraclass correlation coefficients (ICC) and coefficients of variation (COV) and test-retest repeatability as visualized by Bland-Altman analysis. Secondary outcome measures were IOP in sitting (GAT, ICP and TPA) and in reclining (ICP and TPA) position. Results: Mean IOP measured by GAT was 14.9±3.5 mmHg. Mean IOP measured by ICP was 15.6±3.1 mmHg (with TPA 14.8±2.7 mmHg) in sitting and 16.5±3.5 mmHg (with TPA 17.0±3.0 mmHg) in reclining positions. COVs ranged from 2.9% (GAT) to 6.9% (ICP reclining) and ICCs from 0.819 (ICP reclining) to 0.972 (GAT). Conclusions: Repeatability is good with all three devices. GAT has higher repeatability compared to the two tested hand-held devices with lowest COVs and highest ICCs. IOP was higher in the reclining compared to the sitting position. Trial registration: The study was registered to the Clinical Trials Register of the US National Institute of Health, NCT01325324.Keywords
This publication has 34 references indexed in Scilit:
- IOP Variations in the Sitting and Supine PositionsJournal of Glaucoma, 2010
- Dynamic Contour Tonometry in Eyes After Penetrating KeratoplastyCornea, 2009
- Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplastiesBritish Journal of Ophthalmology, 2007
- Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thicknessBritish Journal of Ophthalmology, 2006
- Intraocular Pressure and Ocular Pulse Amplitude Comparisons in Different Types of Glaucoma Using Dynamic Contour TonometryCurrent Eye Research, 2006
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Dynamic TonometryOphthalmologica, 1967
- Automatic Tonometer with Exact Theory: Various Biological ApplicationsScience, 1960
- FAST, AUTOMATIC, ELECTRONIC TONOMETERS BASED ON AN EXACT THEORYActa Ophthalmologica, 1959
- Über ApplanationstonometrieOphthalmologica, 1957