In Vitro and In Vivo Validation Tests for Total Artificial Heart
Authors:
- 1 January 1994
- journal article
- Published by Wiley in Artificial Organs
- Vol. 18 (1) , 54-72
Abstract
Properly planned in vitro and in vivo studies are mandatory to validate blood pump performance. However, standard procedures for in vitro and in vivo tests of total artificial heart (TAH) performance still do not exist. At Baylor College of Medicine, standard tests procedures for performance validation of the totally implantable TAH have been established. In this paper, these in vitro tests protocols (performance mapping tests, accelerated endurance test, hemolysis test, flow visualization tests, etc.) are described as well as in vivo test procedures (TAH implantation procedure, including surgical technique, postoperative management and tests, etc.). In addition, pathological protocols are presented. In our experience, these procedures are very simple, easy to perform, and inexpensive. These protocols are proposed as standard in vitro and in vivo tests for a TAH.Keywords
This publication has 9 references indexed in Scilit:
- The Baylor-ABI Electromechanical Total Artificial HeartASAIO Journal, 1993
- Left and Right Pump Output Control in One‐Piece Electromechanical Total Artificial HeartArtificial Organs, 1993
- Versatile One‐Piece Total Artificial Heart for Bridge to Transplantation or Permanent Heart ReplacementArtificial Organs, 1992
- Baylor Multipurpose Circulatory Support System for Short- to Long-term UseASAIO Journal, 1992
- Is A Totally Implantable Artificial Heart Realistic?Artificial Organs, 1992
- Characterization of Rehydrated Gelatin GelsArtificial Organs, 1991
- Use of the Anaerobic Threshold for Evaluating Various Total Artificial Heart Control Algorithms in CalvesArtificial Organs, 1985
- Anaerobic threshold and respiratory gas exchange during exercise.Journal of Applied Physiology, 1973
- Detecting the threshold of anaerobic metabolism in cardiac patients during exerciseThe American Journal of Cardiology, 1964