Feasibility Study on the Use of Coded Aperture Mask in SPECT for Boron Dose Imaging in TRR-Based BNCT: Preliminary Study
Authors:
Open Access
- 14 December 2024
- journal article
- Published by SCI AND TECH UNIVERSAL INC in Journal of Nuclear Research and Applications
- Vol. 4 (4) , 11-19
Abstract
In this work, the feasibility study on using a Coded-Aperture Imaging based on a Modified Uniformly Redundant Array coded mask in Single Photon Emission Computed Tomography (SPECT) was investigated for boron dose measurement in Boron Neutron Capture Therapy (BNCT). In this preliminary study, a MURA rank 13 with a 2×2 mosaic configuration, and a CZT detector array mounted in a single mask were investigated using Monte Carlo Simulation Code, MCNPX2.7. This is the first investigation on the use of Coded-Aperture Imaging for BNCT purposes for boron dose measurement during treatment. Also, the Maximum Likelihood-Expectation Maximization method was used for image reconstruction for boron dose estimation in the tissue. The results were examined for the recommended therapeutic neutron beam in the Tehran research reactor. Preliminary results show that SPECT with Modified Uniformly Redundant Array masks can be an efficient and precise tool for boron dose measurement in BNCT, and results show the superiority of this method in comparison to the conventional SEPCT in BNCT.Keywords
This publication has 15 references indexed in Scilit:
- Conceptual design of prompt gamma neutron activation analysis facility at Tehran Research Reactor for BNCT applicationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019
- Gamma-ray modulation properties of tungsten coded apertures for a novel mixed-field imaging systemJournal of Instrumentation, 2019
- Coded-aperture imaging systems: Past, present and future development – A reviewRadiation Measurements, 2016
- Design and construction of a thermal neutron beam for BNCT at Tehran Research ReactorApplied Radiation and Isotopes, 2014
- Compact hybrid gamma camera with a coded aperture for investigation of nuclear materialsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- Design of an epithermal neutron beam for BNCT in thermal column of Tehran research reactorAnnals of Nuclear Energy, 2014
- Distributed MLEM: An Iterative Tomographic Image Reconstruction Algorithm for Distributed Memory ArchitecturesIEEE Transactions on Medical Imaging, 2013
- Development of a thick CdTe detector for BNCT–SPECTApplied Radiation and Isotopes, 2011
- An improved method for the computation of the Moore–Penrose inverse matrixApplied Mathematics and Computation, 2011
- Experimental feasibility studies on a SPECT tomograph for BNCT dosimetryApplied Radiation and Isotopes, 2009