Robotic control of a traditional flexible endoscope for therapy
- 21 April 2013
- journal article
- Published by Springer Nature in Journal of Robotic Surgery
- Vol. 7 (3) , 227-234
Abstract
In therapeutic flexible endoscopy a team of physician and assistant(s) is required to control all independent translations and rotations of the flexible endoscope and its instruments. As a consequence the physician lacks valuable force feedback information on tissue interaction, communication errors easily occur, and procedures are not cost-effective. Current tools are not suitable for performing therapeutic procedures in an intuitive and user-friendly way by one person. A shift from more invasive surgical procedures that require external incisions to endoluminal procedures that use the natural body openings could be expected if enabling techniques were available. This paper describes the design and evaluation of a robotic system which interacts with traditional flexible endoscopes to perform therapeutic procedures that require advanced maneuverability. The physician uses one multi-degree-of-freedom input device to control camera steering as well as shaft manipulation of the motorized flexible endoscope, while the other hand is able to manipulate instruments. We identified critical use aspects that need to be addressed in the robotic setup. A proof-of-principle setup was built and evaluated to judge the usability of our system. Results show that robotic endoscope control increases efficiency and satisfaction. Participants valued its intuitiveness, its accuracy, the feeling of being in control, and its single-person setup. Future work will concentrate on the design of a system that is fully functional and takes safety, cleanability, and easy positioning close to the patient into account.Keywords
This publication has 13 references indexed in Scilit:
- Design and evaluation of robotic steering of a flexible endoscope2012
- A technical review of flexible endoscopic multitasking platformsInternational Journal of Surgery, 2012
- High Cecal Intubation Rates With a New Computer-Assisted Colonoscope: A Feasibility StudyAmerican Journal of Gastroenterology, 2011
- Development of a novel endoscopic manipulation system: The Endoscopic operation robotWorld Journal of Gastrointestinal Endoscopy, 2011
- Polypectomy: Basic Principles2009
- The Future of NOTES Instrumentation: Flexible Robotics and In Vivo MinirobotsJournal of Endourology, 2009
- Surgery Without ScarsArchives of Surgery, 2007
- Endoluminal and transluminal surgery: current status and future possibilitiesSurgical Endoscopy, 2006
- ASGE/SAGES Working Group on Natural Orifice Translumenal Endoscopic SurgerySurgical Endoscopy, 2006
- Bimanual coordination and interhemispheric interactionActa Psychologica, 2002