Validation of the pulse decomposition analysis algorithm using central arterial blood pressure
Open Access
- 1 January 2014
- journal article
- research article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 13 (1) , 96
Abstract
There is a significant need for continuous noninvasive blood pressure (cNIBP) monitoring, especially for anesthetized surgery and ICU recovery. cNIBP systems could lower costs and expand the use of continuous blood pressure monitoring, lowering risk and improving outcomes. The test system examined here is the CareTaker® and a pulse contour analysis algorithm, Pulse Decomposition Analysis (PDA). PDA’s premise is that the peripheral arterial pressure pulse is a superposition of five individual component pressure pulses that are due to the left ventricular ejection and reflections and re-reflections from only two reflection sites within the central arteries. The hypothesis examined here is that the model’s principal parameters P2P1 and T13 can be correlated with, respectively, systolic and pulse pressures. Central arterial blood pressures of patients (38 m/25 f, mean age: 62.7 y, SD: 11.5 y, mean height: 172.3 cm, SD: 9.7 cm, mean weight: 86.8 kg, SD: 20.1 kg) undergoing cardiac catheterization were monitored using central line catheters while the PDA parameters were extracted from the arterial pulse signal obtained non-invasively using CareTaker system. Qualitative validation of the model was achieved with the direct observation of the five component pressure pulses in the central arteries using central line catheters. Statistically significant correlations between P2P1 and systole and T13 and pulse pressure were established (systole: R square: 0.92 (p < 0.0001), diastole: R square: 0.78 (p < 0.0001). Bland-Altman comparisons between blood pressures obtained through the conversion of PDA parameters to blood pressures of non-invasively obtained pulse signatures with catheter-obtained blood pressures fell within the trend guidelines of the Association for the Advancement of Medical Instrumentation SP-10 standard (standard deviation: 8 mmHg(systole: 5.87 mmHg, diastole: 5.69 mmHg)). The results indicate that arterial blood pressure can be accurately measured and tracked noninvasively and continuously using the CareTaker system and the PDA algorithm. The results further support the physical model that all of the features of the pressure pulse envelope, whether in the central arteries or in the arterial periphery, can be explained by the interaction of the left ventricular ejection pressure pulse with two centrally located reflection sites.Keywords
This publication has 27 references indexed in Scilit:
- A Reverse Dipping Pattern Predicts Cardiovascular Mortality In a Clinical CohortJournal of Korean Medical Science, 2013
- Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®Anesthesiology, 2012
- Brachial Blood Pressure–Independent Relations Between Radial Late Systolic Shoulder-Derived Aortic Pressures and Target Organ ChangesHypertension, 2012
- Pulse Decomposition Analysis of the digital arterial pulse during hemorrhage simulationNonlinear Biomedical Physics, 2011
- Noninvasive Measurement of Central Vascular Pressures With Arterial Tonometry: Clinical Revival of the Pulse Pressure Waveform?Mayo Clinic Proceedings, 2010
- The Effect of Measuring Ambulatory Blood Pressure on Nighttime Sleep and Daytime Activity—Implications for DippingClinical Journal of the American Society of Nephrology, 2010
- A Comparison of Noninvasive Blood Pressure Measurement on the Wrist with Invasive Arterial Blood Pressure Monitoring in Patients Undergoing Bariatric SurgeryObesity Surgery, 2008
- Force plate monitoring of human hemodynamicsNonlinear Biomedical Physics, 2008
- Superiority of Ambulatory Over Clinic Blood Pressure Measurement in Predicting MortalityHypertension, 2005
- Regional wave travel and reflections along the human aorta: a study with six simultaneous micromanometric pressures.Circulation, 1985