Sub-pixel energy-weighting techniques for metallic contaminant highlighting in a pharmaceutical hard capsule using a Timepix3 CdZnTe hybrid pixel detector
Authors:
Open Access
- 1 October 2022
- journal article
- research article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 17 (10) , P10030
Abstract
The presented work describes the capabilities of a state-of-the-art spectroscopic photon-counting hybrid pixel detector for use in non-destructive testing, as a contaminant detector with potential uses in the food and pharmaceutical industries. A pharmaceutical hard capsule containing vitamin powder, contaminated with Steel (ST) and Tungsten (W) was prepared under controlled conditions. Both contaminants are distinguished from the powder by combining sub-pixel imaging with spectroscopic imaging analysis, made possible by the use of a 2 mm CdZnTe sensor coupled to a Timepix3 ASIC, operated in Time-over-Threshold and Time-of-Arrival mode. The sub-pixelization technique achieves a spatial resolution of 18.3 μm, with a maximum photon flux of 2.5× 106 photons/s· cm2. For the spectral analysis technique implemented, each energy bin image is energy-weighted, with weights optimally defined by considering the bin contrast and noise behavior, and combined to form an integrated-spectrum enhanced image with increased contaminant visibility. Contrast-to-Noise Ratio (CNR) for each contaminant type was used for performance assessment of the approach. A CNR increase of 106% above the reference unweighted-spectrum image was achieved for the ST contaminant, while the denser W contaminant showed a CNR increase of 80%. This means that the presented spectroscopic imaging technique is feasible for implementation in material visibility enhancement.Keywords
This publication has 28 references indexed in Scilit:
- Timepix3: a 65K channel hybrid pixel readout chip with simultaneous ToA/ToT and sparse readoutJournal of Instrumentation, 2014
- The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imagingJournal of Instrumentation, 2013
- Optimal “image‐based” weighting for energy‐resolved CTMedical Physics, 2009
- Imaging theory for X-ray pixel detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Charge-sharing observations with a CdTe pixel detector irradiated with a57Co sourceNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Digital detectors for mammography: the technical challengesEuropean Radiology, 2004
- Atomization process for metal powderMaterials Science and Engineering: A, 2004
- Direct-conversion flat-panel X-ray image sensors for digital radiographyProceedings of the IEEE, 2002
- Detective quantum efficiency dependence on x‐ray energy weighting in mammographyMedical Physics, 1999
- Flat-field correction technique for digital detectors1998