We demonstrate the use of two ultrafast fiber laser systems locked together at identical repetition rates of 100 MHz to achieve a timing resolution below 300 fs for pump-probe experiments. By sweeping the set-point of the locking electronics, we scan the time delay between the individual pulse trains by 800 ps. This scanning technique requires only sub-micrometer mechanical motion. Since the temporal scan range is determined electronically, the acquisition can be limited to regions where meaningful physical data is recorded. We discuss how our technique can approach asynchronous optical sampling based on GHz repetition rate lasers in terms of data collection efficiency while offering a number of practical advantages.