Bone mineral imaged in vivo by 31P solid state MRI of human wrists

Authors:

Abstract
Purpose: To implement solid state 31P MRI (31P SMRI) in a clinical scanner to visualize bone mineral.Materials and Methods: Wrists of seven healthy volunteers were scanned. A quadrature wrist 31P transmit/receive coil provided strong B1 and good signal‐to‐noise ratio (SNR). A 1H‐31P frequency converter was constructed to enable detection of the 31P signal by means of the 1H channel. Data points lost in the receiver dead time were recovered by a second acquisition with longer dwell time and lower gradient strength.Results: Three‐dimensional 31P images, showing only bone mineral of the wrist, were obtained with a clinical 3 Tesla (T) scanner. In the best overall case an image with isotropic resolution of ∼5.1 mm and SNR of 30 was obtained in 37 min. 31P NMR properties (resonance line width 2 kHz and T1 17–19 s) of in vivo human bone mineral were measured.Conclusion: In vivo 31P SMRI visualization of human wrist bone mineral with a clinical MR scanner is feasible with suitable modifications to circumvent the scanners' limitations in reception of short‐T2 signals. Frequency conversion methodology is useful for implementing 31P SMRI measurements on scanners which do not have multinuclear capability or for which the multinuclear receiver dead time is excessive. J. Magn. Reson. Imaging 2011;.

This publication has 13 references indexed in Scilit: