Corneal mechanical sensitivity measurement using a staircase technique
- 26 April 2005
- journal article
- research article
- Published by Springer Nature in Ophthalmic and Physiological Optics
- Vol. 25 (3) , 246-253
Abstract
Purpose: To determine the repeatability of an unequal staircase technique (Garcia‐Perez Staircase, GPS) to measure corneal mechanical threshold using the CRCERT‐Belmonte Aesthesiometer, and to compare this with a previously‐reported psychophysical method (Method of Constant Stimuli, MOCS). Methods: The GPS, utilising unequal ascending and descending steps, was used to obtain a threshold measurement from the mean of six response reversals. Repeatability was determined for the GPS and MOCS methods ( n = 14), and threshold results obtained with each method were compared ( n = 10). Results: The GPS (65.0 ± 16.9 mL min −1 ; CoR ± 18.3 mL min −1 ) method was more repeatable than the MOCS (64.0 ± 15.7 mL min −1 ; CoR ± 37.3 mL min −1 ) and the absolute values obtained with the two methods were not significantly different. Conclusions: Although each method gave equivalent threshold results, the GPS method was more repeatable and quicker to apply and hence should reduce the influence of patient fatigue and help to minimise possible carry‐over effects.Keywords
This publication has 17 references indexed in Scilit:
- Corneal and conjunctival sensitivity to air stimuliBritish Journal of Ophthalmology, 2004
- A Noncontact Pneumatic Esthesiometer for Measurement of Ocular SensitivityCornea, 1999
- A new non‐contact corneal aesthesiometer (NCCA)Ophthalmic and Physiological Optics, 1996
- Variations in corneal sensitivity with hydrogel contact lensesActa Ophthalmologica, 1994
- Response of sensory units with unmyelinated fibres to mechanical, thermal and chemical stimulation of the cat's cornea.The Journal of Physiology, 1993
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Measurement of Corneal Sensitivity and Thickness with PMMA and Gas-Permeable Contact LensesOptometry and Vision Science, 1979
- Effect of Long-term Wear of Hard Contact Lenses on Corneal SensitivityArchives of Ophthalmology (1950), 1978
- EFFECT OF HARD CONTACT LENSES ON CORNEAL SENSITIVITY AND THICKNESSActa Ophthalmologica, 1975
- Diurnal variation of corneal sensitivity.British Journal of Ophthalmology, 1972