Clinical utility of Gafchromic film in an MRI-guided linear accelerator
Open Access
- 26 June 2021
- journal article
- research article
- Published by Springer Nature in Radiation Oncology
- Vol. 16 (1) , 1-12
Abstract
Background: The purpose of this study is to comprehensively evaluate the suitability of Gafchromic EBT3 and EBT-XD film for dosimetric quality assurance in 0.35 T MR-guided radiotherapy. Methods: A 0.35 T magnetic field strength was utilized to evaluate magnetic field effects on EBT3 and EBT-XD Gafchromic films by studying the effect of film exposure time within the magnetic field using two timing sequences and film not exposed to MR, the effect of magnetic field exposure on the crystalline structure of the film, and the effect of orientation of the film with respect to the bore within the magnetic field. The orientation of the monomer crystal was qualitatively evaluated using scanning electron microscopy (SEM) compared to unirradiated film. Additionally, dosimetric impact was evaluated through measurements of a series of open field irradiations (0.83 × 0.83-cm2 to 19.92 × 19.92-cm2) and patient specific quality assurance measurements. Open fields were compared to planned dose and an independent dosimeter. Film dosimetry was applied to twenty conventional and twenty stereotactic body radiotherapy (SBRT) patient specific quality assurance cases. Results: No visual changes in crystal orientation were observed in any evaluated SEM images nor were any optical density differences observed between films irradiated inside or outside the magnetic field for both EBT3 and EBT-XD film. At small field sizes, the average difference along dose profiles measured in film compared to the same points measured using an independent dosimeter and to predicted treatment planning system values was 1.23% and 1.56%, respectively. For large field sizes, the average differences were 1.91% and 1.21%, respectively. In open field tests, the average gamma pass rates were 99.8% and 97.2%, for 3%/3 mm and 3%/1 mm, respectively. The median (interquartile range) 3%/3 mm gamma pass rates in conventional QA cases were 98.4% (96.3 to 99.2%), and 3%/1 mm in SBRT QA cases were 95.8% (95.0 to 97.3%). Conclusions: MR exposure at 0.35 T had negligible effects on EBT3 and EBT-XD Gafchromic film. Dosimetric film results were comparable to planned dose, ion chamber and diode measurements.Keywords
Funding Information
- American Cancer Society (RSG-15-137-01-CCE)
This publication has 29 references indexed in Scilit:
- Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetryMedical Physics, 2015
- A protocol for EBT3 radiochromic film dosimetry using reflection scanningMedical Physics, 2014
- The ViewRay System: Magnetic Resonance–Guided and Controlled RadiotherapySeminars in Radiation Oncology, 2014
- Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetryAustralasian Physics & Engineering Sciences in Medicine, 2013
- Radiochromic Film Dosimetry and its Applications in RadiotherapyAIP Conference Proceedings, 2011
- Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurancePhysics in Medicine & Biology, 2007
- Overview of image-guided radiation therapyMedical Dosimetry, 2006
- Magnetic field polymerisationProgress in Polymer Science, 2000
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- 3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC)Medical Physics, 1996