Abstract
Monte Carlo simulations using the code PENELOPE have been performed to test a simplified model of the source channel geometry of the Leksell GammaKnife. The characteristics of the radiation passing through the treatment helmets are analysed in detail. We have found that only primary particles emitted from the source with polar angles smaller than 3 degrees with respect to the beam axis are relevant for the dosimetry of the Gamma Knife. The photon trajectories reaching the output helmet collimators at (x, v, z = 236 mm) show strong correlations between rho = (x2 + y2)(1/2) and their polar angle theta, on one side, and between tan(-1)(y/x) and their azimuthal angle phi, on the other. This enables us to propose a simplified model which treats the full source channel as a mathematical collimator. This simplified model produces doses in good agreement with those found for the full geometry. In the region of maximal dose, the relative differences between both calculations are within 3%, for the 18 and 14 mm helmets, and 10%, for the 8 and 4 mm ones. Besides, the simplified model permits a strong reduction (larger than a factor 15) in the computational time.Keywords
All Related Versions
This publication has 11 references indexed in Scilit:
- Monte Carlo simulation of the Leksell Gamma Knife$reg$: I. Source modelling and calculations in homogeneous mediaPhysics in Medicine & Biology, 2002
- Monte Carlo simulation of electron beams from an accelerator head using PENELOPEPhysics in Medicine & Biology, 2001
- Dose distributions at extreme irradiation depths of gamma knife radiosurgery: EGS4 Monte Carlo calculationsApplied Radiation and Isotopes, 2001
- Stereotactic dose planning system used in Leksell Gamma Knife model-B: EGS4 Monte Carlo versus GafChromic films MD-55Applied Radiation and Isotopes, 2000
- Monte Carlo calculated output factors of a Leksell Gamma Knife unitPhysics in Medicine & Biology, 1999
- Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unitMedical Physics, 1999
- Simulation of dose distribution irradiation by the Leksell Gamma UnitPhysics in Medicine & Biology, 1999
- Monte Carlo treatment planning for stereotactic radiosurgeryRadiotherapy and Oncology, 1998
- Monte Carlo calculation of single‐beam dose profiles used in a gamma knife treatment planning systemMedical Physics, 1998
- Physics of gamma knife approach on convergent beams in stereotactic radiosurgeryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1990