Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study
Authors:
- 4 May 2011
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 28 (2) , 415-423
Abstract
To investigate the image quality and dose performance of 80 kV high-pitch spiral (HPS) coronary CT angiography (CCTA). 106 patients consecutively enrolled into prospectively ECG-triggering HPS CCTA (pitch = 3.4) exam using kV/ref. mAs = 80/400, 100/370, and 120/370 when patient BMI was ≤22.5 (n = 40), between 22.5 and 27.5 (n = 53) and >27.5 kg/m² (n = 13). Image quality was assessed per-segment by two observers independently using a 4-point scale (1—excellent, 4—non-diagnosable). Image noise and signal-to-noise ratio (SNR), contrast-to-noise ratio were measured. Diagnostic image quality was obtained in 503 of 507, 687 of 693, 164 of 167 coronary segments in 80, 100, 120 kV groups without significant difference (P = 0.482). The proportions of segments with score 1–4 were not significantly different among three kV groups (all P > 0.05). Image noise were significantly higher in 80 kV group than 100 and 120 groups (P < 0.001), while SNR was not (P = 0.097). The effective dose of 80 kV group (0.36 ± 0.03 mSv) was significantly lower than that of 100 kV group (0.86 ± 0.08 mSv) and 120 kV group (1.77 ± 0.18 mSv). The mean ± SD of HR in all patients was 54.8 ± 5.1 bpm. 80 kV HPS CCTA is feasible for patient with BMI ≤ 22.5 kg/m² which can save 58% dose than 100 kV group, while maintain diagnosable image quality.Keywords
This publication has 23 references indexed in Scilit:
- Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experienceThe International Journal of Cardiovascular Imaging, 2010
- Dual-Source versus 64-Section CT Coronary Angiography at Lower Heart Rates: Comparison of Accuracy and Radiation DoseRadiology, 2009
- Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experienceEuropean Radiology, 2009
- Radiation dose exposure of computed tomography coronary angiography: comparison of dual-source, 16-slice and 64-slice CTHeart, 2009
- Image Quality in a Low Radiation Exposure Protocol for Retrospectively ECG-Gated Coronary CT AngiographyAmerican Journal of Roentgenology, 2009
- High-pitch spiral acquisition: A new scan mode for coronary CT angiographyJournal of Cardiovascular Computed Tomography, 2009
- Radiation dose of cardiac dual-source CT: The effect of tailoring the protocol to patient-specific parametersEuropean Journal of Radiology, 2008
- Dual-Source CT in Step-and-Shoot Mode: Noninvasive Coronary Angiography with Low Radiation Dose1Radiology, 2008
- Estimating Risk of Cancer Associated With Radiation Exposure From 64-Slice Computed Tomography Coronary AngiographyJAMA, 2007
- MDCT of the Coronary Arteries: Feasibility of Low-Dose CT with ECG-Pulsed Tube Current Modulation to Reduce Radiation DoseAmerican Journal of Roentgenology, 2006