Ascorbic acid determination with an automated enzymatic procedure
Authors:
Open Access
- 1 January 1997
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 43 (1) , 154-157
Abstract
Given the widespread interest in antioxidant nutrients, we have developed a new procedure that will permit the automated determination of plasma ascorbic acid (AA) concentration with a Roche Fara centrifugal analyzer. After the deproteinization of plasma with metaphosphoric acid, AA is oxidized to dehydroascorbic acid by AA oxidase. The product is coupled to o-phenylenediamine to produce a chromophore for which the absorbance is measured at 340 nm. The procedure allows much faster throughput than conventional HPLC methods while yielding results that correlate well and provide improved precision. Linearity extends beyond the reference range of 26.1–84.6 μmol/L, and severe hemolysis is the only interference identified.Keywords
This publication has 5 references indexed in Scilit:
- Measurement of ascorbic acid in human plasma and serum: stability, intralaboratory repeatability, and interlaboratory reproducibilityClinical Chemistry, 1996
- Total vitamin C, ascorbic acid, and dehydroascorbic acid concentrations in plasma of critically ill patientsThe American Journal of Clinical Nutrition, 1996
- Ascorbic acid and dehydroascorbic acid analyses in biological samplesAnalytical Biochemistry, 1992
- Ascorbic acid status: biochemical and clinical considerationsThe American Journal of Clinical Nutrition, 1988
- Ascorbic acid in lymphocytes: cell preparation and liquid-chromatographic assay.Clinical Chemistry, 1982