Verification of the plan dosimetry for high dose rate brachytherapy using metal–oxide–semiconductor field effect transistor detectors
Authors:
- Zhen‐Yu QiZQZhen‐Yu Qi
- Xiao‐Wu DengXDXiao‐Wu Deng
- Shao‐Min HuangSHShao‐Min Huang
- Jie LuJLJie Lu
- Michael LerchMLMichael Lerch
- Dean CutajarDCDean Cutajar
- Anatoly RosenfeldARAnatoly Rosenfeld
- 11 May 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (6Part1) , 2007-2013
Abstract
The feasibility of a recently designed metal–oxide–semiconductor field effect transistor (MOSFET) dosimetry system for dose verification of high dose rate (HDR) brachytherapy treatment planning was investigated. MOSFET detectors were calibrated with a NE‐2571 Farmer‐type ionization chamber in water. Key characteristics of the MOSFET detectors, such as the energy dependence, that will affect phantom measurements with HDR sources were measured. The MOSFET detector was then applied to verify the dosimetric accuracy of HDR brachytherapy treatments in a custom‐made water phantom. Three MOSFET detectors were calibrated independently, with the calibration factors ranging from . A distance dependent energy response was observed, significant within from the source. The new MOSFET detector has a good reproducibility , small angular effect , and good dose linearity . It was observed that the MOSFET detectors had a linear response to dose until the threshold voltage reached approximately for source measurements. Further comparison of phantom measurements using MOSFET detectors with dose calculations by a commercial treatment planning system for computed tomography‐based brachytherapy treatment plans showed that the mean relative deviation was for dose points away from the source and for dose points located away. The percentage deviations between the measured doses and the planned doses were below 5% for all the measurements. The MOSFET detector, with its advantages of small physical size and ease of use, is a reliable tool for quality assurance of HDR brachytherapy. The phantom verification method described here is universal and can be applied to other HDR brachytherapy treatments.Keywords
This publication has 23 references indexed in Scilit:
- Favorable clinical outcomes of three-dimensional computer-optimized high-dose-rate prostate brachytherapy in the management of localized prostate cancerBrachytherapy, 2006
- Optimization of dose distribution for HDR brachytherapy of the prostate using Attraction-Repulsion ModelInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Customized conformal high-dose-rate brachytherapy boost for limited-volume nasopharyngeal cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- The use of linear programming in optimization of HDR implant dose distributionsMedical Physics, 2003
- Inverse planning anatomy‐based dose optimization for HDR‐brachytherapy of the prostate using fast simulated annealing algorithm and dedicated objective functionMedical Physics, 2001
- Quality control in interstitial brachytherapy of the breast using pulsed dose rate: treatment planning and dose delivery with an Ir-192 afterloading systemRadiotherapy and Oncology, 2001
- Treatment planning considerations and quality assurance for CT-guided transischiorectal implantation of the prostateMedical Physics, 1999
- Absorbed dose determination at short distance from60Co and192Ir brachytherapy sourcesPhysics in Medicine & Biology, 1998
- High dose-rate brachytherapy treatment delivery: Report of the AAPM Radiation Therapy Committee Task Group No. 59Medical Physics, 1998
- Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56Medical Physics, 1997