Turbidimetry of rheumatoid factor in serum with a centrifugal analyzer.
Authors:
- L BorqueLBL Borque
- M YagoMYM Yago
- C MarCMC Mar
- C RodriguezCRC Rodriguez
Open Access
- 1 January 1986
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 32 (1) , 124-129
Abstract
We describe the simple, rapid turbidimetry of IgM rheumatoid factor in human serum by use of the Cobas-Bio centrifugal analyzer. Heat-aggregated human IgG is used as the antigen. The immunoturbidimetric reaction is monitored at 340 nm for 300 s, and the changes in absorbance after the antigen is added are used to prepare the standard curve. Test results are calculated from the stored curve and reported in int. units/mL, based on comparison with the WHO reference serum for rheumatoid factor. There is no interference from bilirubin (up to 340 mumol/L) or hemoglobin (up to 5600 mg/L). Serum samples with a triglyceride concentration greater than 2.20 mmol/L must be cleared of lipids before analysis. The standard curve is linear from 30 to 500 int. units/mL. Precision, accuracy, linearity, and sensitivity are quite acceptable. The CV was generally less than 5% for different concentrations of rheumatoid factor. Results agree well with those by a rate-nephelometric procedure on the Beckman ICS system (rs = 0.932). However, both correlate poorly with a modified classical Waaler-Rose test. Of 47 patients with rheumatoid arthritis, 34 had IgM rheumatoid factor in their serum, but the measured value did not reflect the activity of the disease.Keywords
This publication has 21 references indexed in Scilit:
- A sensitive radioimmunoassay for quantitation of igm rheumatoid factorArthritis & Rheumatism, 1980
- Rate nephelometric measurement of rheumatoid factor in serum.Clinical Chemistry, 1979
- Performance of Latex-fixation Kits Used for Serologic Diagnosis of Rheumatoid Factor in Rheumatoid Arthritis SeraAmerican Journal of Clinical Pathology, 1979
- Quantitation of Rheumatoid Factor Activity by NephelometryAmerican Journal of Clinical Pathology, 1979
- Kinetic immunoturbidimetry: the estimation of albuminClinica Chimica Acta, 1979
- Prognosis in rheumatoid arthritisThe American Journal of Medicine, 1979
- Rheumatoid factor: Its nature, specificity, and production in rheumatoid arthritisClinical Immunology and Immunopathology, 1976
- Immune Reactions in Polysaccharide Media. I. The Effect of Dextran on the Reaction between Iodine-125 Labelled Human Serum Albumin and gammaG-Globulin from Rabbit Anti-albumin Sera.Acta Chemica Scandinavica, 1966
- Differential Agglutination of Normal and Sensitized Sheep Erythrocytes by Sera of Patients with Rheumatoid ArthritisExperimental Biology and Medicine, 1948
- ON THE OCCURRENCE OF A FACTOR IN HUMAN SERUM ACTIVATING THE SPECIFIC AGGLUTINATION OF SHEEP BLOOD CORPUSCLES.Acta Pathologica Microbiologica Scandinavica, 1940