Biocompatible Multifunctional Black‐Silicon for Implantable Intraocular Sensor

Authors:

Abstract
Multifunctional black‐silicon (b‐Si) integrated on the surface of an implantable intraocular pressure sensor significantly improves sensor performance and reliability in six‐month in vivo studies. The antireflective properties of b‐Si triples the signal‐to‐noise ratio and increases the optical readout distance to a clinically viable 12 cm. Tissue growth and inflammation response on the sensor is suppressed demonstrating desirable anti‐biofouling properties.
Funding Information
  • National Institutes of Health (EY024582)

This publication has 34 references indexed in Scilit: