A new high‐aperture glycerol immersion objective lens and its application to 3D‐fluorescence microscopy
- 7 May 2002
- journal article
- research article
- Published by Wiley in Journal of Microscopy
- Vol. 206 (2) , 146-151
Abstract
Summary: High‐resolution light microscopy of glycerol‐mounted biological specimens is performed almost exclusively with oil immersion lenses. The reason is that the index of refraction of the oil and the cover slip of ~1.51 is close to that of ~1.45 of the glycerol mountant, so that refractive index mismatch‐induced spherical aberrations are tolerable to some extent. Here we report the application of novel cover glass‐corrected glycerol immersion lenses of high numerical aperture (NA) and the avoidance of these aberrations. The new lenses feature a semi‐aperture angle of 68.5°, which is slightly larger than that of the diffraction‐limited 1.4 NA oil immersion lenses. The glycerol lenses are corrected for a quartz cover glass of 220 µm thickness and for a 80% glycerol‐water immersion solution. Featuring an aberration correction collar, the lens can adapt to glycerol concentrations ranging between 72% and 88%, to slight variations of the temperature, and to the cover glass thickness. As the refractive index mismatch‐induced aberrations are particularly important to quantitative confocal fluorescence microscopy, we investigated the axial sectioning ability and the axial chromatic aberrations in such a microscope as well as the image brightness as a function of the penetration depth. Whereas there is a significant decrease in image brightness associated with oil immersion, this decrease is absent with the glycerol immersion system. In addition, we show directly the compression of the optic axis in the case of oil immersion and its absence in the glycerol system. The unique advantages of these new lenses in high‐resolution microscopy with two coherently used opposing lenses, such as 4 Pi‐microscopy, are discussed.Keywords
This publication has 15 references indexed in Scilit:
- I5M: 3D widefield light microscopy with better than 100 nm axial resolutionJournal of Microscopy, 1999
- Equivalence of the Huygens–Fresnel and Debye approach for the calculation of high aperture point‐spread functions in the presence of refractive index mismatchJournal of Microscopy, 1999
- 4Pi-Confocal Microscopy Provides Three-Dimensional Images of the Microtubule Network with 100- to 150-nm ResolutionJournal of Structural Biology, 1998
- Refractive index mismatch induced intensity and phase variations in fluorescence confocal, multiphoton and 4Pi-microscopyOptics Communications, 1998
- Ultrathin fluorescent layers for monitoring the axial resolution in confocal and two‐photon fluorescence microscopyJournal of Microscopy, 1998
- The role of specimen‐induced spherical aberration in confocal microscopyJournal of Microscopy, 1997
- Far‐field fluorescence microscopy with three‐dimensional resolution in the 100‐nm rangeJournal of Microscopy, 1997
- Aberrations in confocal fluorescence microscopy induced by mismatches in refractive indexJournal of Microscopy, 1993
- Properties of a 4Pi confocal fluorescence microscopeJournal of the Optical Society of America A, 1992
- Electromagnetic diffraction in optical systems, II. Structure of the image field in an aplanatic systemProceedings of the Royal Society of London. A. Mathematical and Physical Sciences, 1959