Novel device for automating exchange transfusions through umbilical venous route in neonates
- 10 January 2023
- journal article
- research article
- Published by Springer Nature in European Journal of Pediatrics
- Vol. 182 (3) , 1229-1238
Abstract
Manually performed double-volume exchange transfusion (DVET) is tedious, error-prone, and may incur the risk of embolism. We aimed to develop a device that automates the DVET procedure performed through the umbilical venous route. We evaluated changes in blood passing through the device during DVET. We developed an electro-mechanical device with accessories (tubing and valve assembly) to perform a complete DVET. It comprises two syringes driven by a common pump that moves back and forth to withdraw aliquots of the patient’s blood and infuse equal volumes of donor blood. In tandem, it draws donor blood from a blood bank bag and pushes the patient blood drawn from the previous cycle into a waste bag, respectively. One-way duckbill valves and a two-way pinch valve ensure the separation of the donor and patient blood. A sensor detects bubbles and clots. A dashboard displays set and measured parameters. We tested the accuracy of the delivered flow rate and volume, electrical safety, embolus detection, and changes in hematological and biochemical values. The delivered flow and volume were within 5% of the set parameters. All electrical safety parameters were within normal limits. The sensor consistently detected microbubbles and clots. There were no clinically significant differences in laboratory parameters between samples drawn directly from the blood bank bag and drawn from the exit port at 80, 100, 120, and 160 s with a fixed aliquot volume. Conclusions: Our prototype of a novel device can safely automate a DVET. Further trials of this device are warranted. What is Known: • Double volume exchange transfusion is often performed manually, but this is time-consuming and error-prone. • Previous attempts at automation were not widely adopted because they involved inserting two catheters and did not have mechanisms to prevent embolism. What is New: • This novel device fully automates double volume exchange transfusions through a single-lumen umbilical venous catheter. • It prevents air and clot embolism and has a screen for input and output parameters and alarms.Keywords
Funding Information
- Department of Science and Technology, Government of India (IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018, IDP/MED/03/2016 dated 7 May 2018)
This publication has 16 references indexed in Scilit:
- Comparison of intra-procedural pain between a novel continuous arteriovenous exchange and conventional pull-push techniques of partial exchange transfusion in neonates: a randomized controlled trialJournal of Perinatology, 2014
- Blood Exchange Transfusion for Infants with Severe Neonatal HyperbilirubinemiaSeminars in Perinatology, 2011
- Rhesus haemolytic disease of the newborn: Postnatal management, associated morbidity and long-term outcomeSeminars in Fetal and Neonatal Medicine, 2008
- Late hyporegenerative anemia in neonates with rhesus hemolytic diseaseJournal of Perinatal Medicine, 1999
- Automated apparatus for exchange transfusionThe Journal of Pediatrics, 1984
- Automated method for exchange transfusion: A new modificationThe Journal of Pediatrics, 1983
- Automated method for exchange transfusion.Archives of Disease in Childhood, 1973
- A continuous isovolumetric exchange transfusion techniqueThe Journal of Pediatrics, 1970
- Simultaneous umbilical arteriovenous exchange transfusion.BMJ, 1966
- Erythroblastosis FetalisNew England Journal of Medicine, 1951