Microbial contamination of mist therapy units on six pediatric wards.

Authors:
  • NPN J Petersen
    N J Petersen
  • LCL A Carson
    L A Carson
  • MFM S Favero
    M S Favero
  • JMJ H Marshall
    J H Marshall
  • WBW W Bond
    W W Bond

  • 1 January 1975
    • journal article
    • research article
Abstract
A study was conducted in six pediatric wards to investigated factors that influence microbial contamination of mist therapy units. Most samples from nebulizer reservoirs were positive for gram-negative bacilli which multiplied rapidly in distilled water and reached levels as high as 3.0 X 10(8) viable microorganisms per ml. High prevalence rates of Pseudomonas aeruginosa and coliform bacteria as well as high levels of gram-negative bacilli were found to be associated with infrequent and inadequate cleaning and disinfection of the nebulizing equipment. In one of the wards exhibiting high rates, a move to new quarters permitted the implementation of improved cleaning and maintenance procedures for mist therapy units. The result was a significant reduction of microbial contamination. This study reemphasizes the role of microbiological surveillance in assuring proper care of nebulizing equipment in the hospital.