Lead(II)-Selective Electrode Based on Phenyl Disulfide.
Authors:
- 1 January 2002
- journal article
- Published by Springer Nature in Analytical Sciences
- Vol. 18 (9) , 987-991
Abstract
A PVC membrane electrode for lead ions based on phenyl disulfide as the membrane carrier was developed. The electrode exhibits a good Nernstian slope of 29.3 +/- 0.7 mV/decade and a linear range of 2.0 x 10(-6)-1.0 x 10(-2) M for Pb(NO3)2. The limit of detection is 1.2 x 10(-6) M. It has a response time of 45 s and can be used for at least fifty days without any divergence in potential. The proposed membrane sensor revealed high selectivity for Pb2+ over a wide variety of other metal ions and could be used in the pH range of 3.5-6.3. The electrode was used as an indicator electrode in potentiometric titration of lead ions.Keywords
This publication has 17 references indexed in Scilit:
- Highly selective electrode for Nickel (II) ion based on 1.5-diphenylthiocarbazone (dithizone)Microchemical Journal, 2001
- A PVC-based capric acid membrane potentiometric sensor for lead(II) ionsSensors and Actuators B: Chemical, 2001
- Chromium(III) ion-selective electrode based on 4-dimethylaminoazobenzeneTalanta, 2001
- Lead(II)-Selective Membrane Electrode Based on TetraphenylporphyrinAnalytical Letters, 2000
- Application of Thiacrown Ethers as Ligands for Lead-Selective ElectrodesAnalytical Sciences, 1999
- General Description of the Simultaneous Response of Potentiometric Ionophore-Based Sensors to Ions of Different ChargeAnalytical Chemistry, 1999
- Lead-selective electrode by using benzyl disulphide as ionophoreAnalytica Chimica Acta, 1999
- Lead(II) ion-selective electrodes based on crown ethersThe Analyst, 1992
- Recent Advances in the Development and Analytical Applications of Biosensing ProbesC R C Critical Reviews in Analytical Chemistry, 1988
- Transition metal ion-selective membrane electrodes based on complexing compounds with heteroatoms. Part I. Complexing compounds containing nitrogen atomsThe Analyst, 1988