Polarized high-brilliance and high-resolution soft x-ray source at ELETTRA: The performance of beamline BACH
- 1 January 2004
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 75 (1) , 31-36
Abstract
BACH, a soft x-ray beamline for polarization-dependent experiments at the Italian synchrotron radiation facility ELETTRA, was recently completed and characterized. Its performance, in terms of energy resolution, flux and polarization, is presented. Based on two APPLE II undulators, BACH covers the energy range between 35 and 1600 eV with the control of the light polarization. The monochromator is equipped with four gratings and allows one to work either in a high resolution or in a high flux mode. After the monochromator, the beamline is split into two branches with different refocusing properties. One is optimized to exploit the performance of the soft x-ray spectrometer (ComIXS) available at the beamline. Resolving powers between 12000 at 90 eV photon energy and 6600 near 867 eV were achieved using the high-resolution gratings and the smallest available slit width For the high-brilliance grating, which works between 290 and 1600 eV, resolving powers between 7000 at 400 eV and 2200 at 867 eV were obtained. The flux in the experimental chamber, measured with the high-resolution gratings for linearly polarized light at the best achievable resolution, ranges between at 125 eV and between 900 and 1250 eV. In circularly polarized mode the flux is two times larger for energies up to 380 eV. A gain of nearly one order of magnitude is obtained for the high-brilliance grating, in accordance with theoretical predictions. Flux beyond was measured up to 1300 eV, and thus over nearly the complete energy range covered by this high-brilliance grating, with a maximum of between 800 and 1100 eV. First results from polarization measurements confirm a polarization above 99.7% for both linearly and circularly polarized modes at low energies. Circular dichroism experiments indicate a circular polarization beyond 90% at the edge near 720 eV.
Keywords
This publication has 16 references indexed in Scilit:
- ComIXS: a compact inelastic x-ray spectrometer2001
- BACH, the beamline for advanced dichroic and scattering experiments at ELETTRAReview of Scientific Instruments, 2001
- Performance of the grating-crystal monochromator of the ALOISA beamline at the Elettra SynchrotronReview of Scientific Instruments, 1999
- A critical comparison of selected 1s and 2p core hole widthsJournal of Electron Spectroscopy and Related Phenomena, 1999
- The gas-phase photoemission beamline at ElettraJournal of Synchrotron Radiation, 1998
- Performance of a varied-space plane-grating monochromator1997
- High-resolution pre-edge structure in the inner-shell ionization threshold region of rare gases Xe, Kr, and ArPhysical Review A, 1996
- Analyses for a planar variably-polarizing undulatorNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1994
- Optimization of soft x-ray monochromators (invited)Review of Scientific Instruments, 1989
- An investigation of the structure near the L2,3edges of argon, the M4,5edges of krypton and the N4,5edges of xenon, using electron impact with high resolutionJournal of Physics B: Atomic and Molecular Physics, 1977