Cardiac MR elastography using reduced‐FOV, single‐shot, spin‐echo EPI
Open Access
- 1 December 2017
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 80 (1) , 231-238
Abstract
Purpose: To implement a reduced field of view (rFOV) technique for cardiac MR elastography (MRE) and to demonstrate the improvement in image quality of both magnitude images and post‐processed MRE stiffness maps compared to the conventional full field of view (full‐FOV) acquisition.Methods: With Institutional Review Board approval, 17 healthy volunteers underwent both full‐FOV and rFOV cardiac MRE scans using 140‐Hz vibrations. Two cardiac radiologists blindly compared the magnitude images and stiffness maps and graded the images based on several image quality attributes using a 5‐point ordinal scale. Fisher's combined probability test was performed to assess the overall evaluation. The octahedral shear strain‐based signal‐to‐noise ratio (OSS‐SNR) and median stiffness over the left ventricular myocardium were also compared.Results: One volunteer was excluded because of an inconsistent imaging resolution during the exam. In the remaining 16 volunteers (9 males, 7 females), the rFOV scans outperformed the full‐FOV scans in terms of subjective image quality and ghosting artifacts in the magnitude images and stiffness maps, as well as the overall preference. The quantitative measurements showed that rFOV had significantly higher OSS‐SNR (median: 1.4 [95% confidence interval (CI): 1.2–1.5] vs. 2.1 [95% CI: 1.8–2.4]), P < 0.05) compared to full‐FOV. Although no significant change was found in the median myocardial stiffness between the 2 scans, we observed a decrease in the stiffness variation within the myocardium from 2.1 kPa (95% CI: [1.9, 2.3]) to 1.9 kPa (95% CI: [1.7, 2.0]) for full‐FOV and rFOV, respectively (P < 0.05) in a subgroup of 7 subjects with ghosting present in the myocardium.Conclusion: This pilot volunteer study demonstrated that rFOV cardiac MRE has the capability to reduce ghosting and to improve image quality in both MRE magnitude images and stiffness maps. Magn Reson Med 80:231–238, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.Keywords
Funding Information
- National Institutes of Health (R01_HL115144)
This publication has 32 references indexed in Scilit:
- Parallel EPI artifact correction (PEAC) for N/2 ghost suppression in neuroimaging applicationsMagnetic Resonance Imaging, 2013
- Prospective Comparison of Cartesian Acquisition with Projection-like Reconstruction Magnetic Resonance Angiography with Computed Tomography Angiography for Evaluation of below-the-Knee RunoffJournal of Vascular and Interventional Radiology, 2013
- An octahedral shear strain-based measure of SNR for 3D MR elastographyPhysics in Medicine & Biology, 2011
- Cardiac Magnetic Resonance ElastographyInvestigative Radiology, 2010
- Magnetic resonance elastography as a method for the assessment of effective myocardial stiffness throughout the cardiac cycleMagnetic Resonance in Medicine, 2010
- High‐resolution diffusion tensor imaging with inner field‐of‐view EPIJournal of Magnetic Resonance Imaging, 2009
- Concomitant field terms for asymmetric gradient coils: Consequences for diffusion, flow, and echo‐planar imagingMagnetic Resonance in Medicine, 2008
- Rapid MR elastography using selective excitationsMagnetic Resonance in Medicine, 2006
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991