High-Order Statistical Blind Deconvolution of Spectroscopic Data with a Gauss—Newton Algorithm
Authors:
- 1 June 2006
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 60 (6) , 692-697
Abstract
The spectroscopic data recorded by a dispersion spectrophotometer are usually degraded by the response function of the instrument. To improve the resolving power, double or triple cascade spectrophotometers and narrow slits have been employed, but the total flux of the radiation available decreases accordingly, resulting in a lower signal-to-noise ratio (SNR) and a longer measurement time. However, the spectral resolution can be improved by mathematically removing the effect of the instrument response function. A high-order statistical Gauss–Newton algorithm is proposed to blindly deconvolve the measured spectroscopic data. The true spectrum and the instrument response function are estimated simultaneously. Experiments on artificial and real measured spectroscopic data demonstrate the feasibility of this method.Keywords
This publication has 3 references indexed in Scilit:
- High-order cumulant-based blind deconvolution of Raman spectraApplied Optics, 2005
- Estimation of spectral slit width and blind deconvolution of spectroscopic data by homomorphic filteringApplied Optics, 1984
- Image Restoration by the Method of Least SquaresJournal of the Optical Society of America, 1967