Scatter fractions in AMBER imaging.
Authors:
- C E FloydCFC E Floyd
- J T DobbinsJDJ T Dobbins
- J Y LoJLJ Y Lo
- C E RavinCRC E Ravin
- 1 December 1990
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 177 (3) , 879-880
Abstract
Images of two phantoms were obtained with use of an advanced multiple-beam equalization radiography system, and scatter fractions were estimated with use of a photostimulable phosphor imaging system. Scatter fractions in the equalized images were lower in the mediastinum-equivalent areas and higher in the lung-equivalent areas, relative to images that were conventionally acquired with use of an antiscatter grid. The differences are attributed to a reduction in incident exposure in the lungs and the presence of cross-scatter between lung and mediastinal regions.Keywords
This publication has 4 references indexed in Scilit:
- Quantitative radiographic imaging using a photostimulable phosphor systemMedical Physics, 1990
- AMBER: a scanning multiple-beam equalization system for chest radiography.Radiology, 1988
- Digital chest radiography: performance evaluation of a prototype unit.Radiology, 1985
- A scanning system for chest radiography with regional exposure control: Practical implementationMedical Physics, 1983