Use of calcium-titrated sodium heparinate for ionized calcium determination on plasma and whole blood (heparinized blood sampler B-129 Radiometer)
Authors:
- 1 January 1991
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 51 (3) , 303-306
Abstract
We report modifications of ionized calcium as observed on specimens collected in a commercially available sampling device (Radiometer B-129). When the sampler is used as recommended, errors are non-significant around 1.25 mmol l-1. There is a +3% increase for low (0.75 mmol l-1) values and a -2% decrease for high (2.50 mmol l-1) values. Incomplete syringe-filling increases these errors.Keywords
This publication has 4 references indexed in Scilit:
- Anticoagulant-induced preanalytical errors in ionized calcium determination on bloodScandinavian Journal of Clinical and Laboratory Investigation, 1989
- Neonatal reference values for ionized calcium, phosphate and magnesium. Selection of reference population by optimality criteriaScandinavian Journal of Clinical and Laboratory Investigation, 1987
- Evaluation of Radiometer ICA1 as a routine instrument for serum ionized calcium and its application for whole blood capillary samples from newborn infantsScandinavian Journal of Clinical and Laboratory Investigation, 1983
- Ionized calcium in normal serum, ultrafiltrates, and whole blood determined by ion-exchange electrodesJournal of Clinical Investigation, 1970