An improved passive shimming approach to design correction iron pieces for high field MRI
Authors:
- 1 December 2020
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 91 (12) , 124105
Abstract
This paper proposes a new passive shimming method to design correction iron pieces for compensating field impurities generated by the main magnets of high field magnetic resonance imaging (MRI) for human body imaging. First, the relationship formula of a magnetic dipole to any field point in space is established. Then, a sensitivity matrix in the form of spherical harmonic expansions can be set up based on the contribution of each shim piece to each field point. Next, an optimization procedure of linear programming is applied to determine the location and thickness of the ferromagnetic shim pieces. This is different from the previous methods that consider that all magnetic moments for each magnetized iron piece are located only at its central position, which may cause some intrinsic errors. This method takes the shim piece’s volume into account and gives more accurate results in a sensitivity matrix by means of integration in the azimuthal and axial directions while the radial (thickness) direction remains constant. Finally, a case study combining the analytical method and a 3D finite element analysis simulation demonstrates that the new approach provides better results in terms of homogeneity. The method presented can also be used to design passive shims for various practical MRI applications.Keywords
Funding Information
- Natural Science Research of Jiangsu Higher Education Institutions of China (19KJB470011)
- Applied Basic Research Program of Changzhou, China (J20200011)
- Strategic Priority Research Program of Chinese Academy of Sciences (XDB25000000)
This publication has 13 references indexed in Scilit:
- Tesseral superconducting shim coil design with quasi-saddle geometry for use in high-field magnet systemReview of Scientific Instruments, 2019
- MRI and MRS of the human brain at magnetic fields of 14 T to 20 T: Technical feasibility, safety, and neuroscience horizonsNeuroImage, 2018
- A novel passive shimming method for the correction of magnetic fields above the patient bed in MRIJournal of Magnetic Resonance, 2015
- High magnetic field science and its application in the United States: A magnetic resonance perspectiveJournal of Magnetic Resonance, 2014
- Passive Shimming by Eliminating Spherical Harmonics Coefficients of all Magnetic Field Components Generated by Correction Iron PiecesIEEE Transactions on Magnetics, 2014
- A Hybrid Field-Harmonics Approach for Passive Shimming Design in MRIIEEE Transactions on Applied Superconductivity, 2011
- Tests of a Prototype for Assessing the Field Homogeneity of the Iseult/Inumac 11.7 T Whole Body MRI MagnetIEEE Transactions on Applied Superconductivity, 2010
- Passive Shim Design and a Shimming Approach for Biplanar Permanent Magnetic Resonance Imaging MagnetsIEEE Transactions on Magnetics, 2008
- Three-Dimensional Optimization of Correction Iron Pieces for Open High Field MRI SystemIEEE Transactions on Applied Superconductivity, 2004
- A novel target-field method for finite-length magnetic resonance shim coils: I. Zonal shimsJournal of Physics D: Applied Physics, 2001