Real-time controller for foot-drop correction by using surface electromyography sensor
- 9 January 2013
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 227 (4) , 373-383
Abstract
Foot drop is a disease caused mainly by muscle paralysis, which incapacitates the nerves generating the impulses that control feet in a heel strike. The incapacity may stem from lesions that affect the brain, the spinal cord, or peripheral nerves. The foot becomes dorsiflexed, affecting normal walking. A design and analysis of a controller for such legs is the subject of this article. Surface electromyography electrodes are connected to the skin surface of the human muscle and work on the mechanics of human muscle contraction. The design uses real surface electromyography signals for estimation of the joint angles. Various-speed flexions and extensions of the leg were analyzed. The two phases of the design began with surface electromyography of real human leg electromyography signal, which was subsequently filtered, amplified, and normalized to the maximum amplitude. Parameters extracted from the surface electromyography signal were then used to train an artificial neural network for prediction of the joint angle. The artificial neural network design included various-speed identification of the electromyography signal and estimation of the angles of the knee and ankle joints by a recognition process that depended on the parameters of the real surface electromyography signal measured through real movements. The second phase used artificial neural network estimation of the control signal, for calculation of the electromyography signal to be stimulated for the leg muscle to move the ankle joint. Satisfactory simulation (MATLAB/Simulink version 2012a) and implementation results verified the design feasibility.Keywords
This publication has 31 references indexed in Scilit:
- A Novel Method for Measuring In-Shoe Navicular Drop during GaitSensors, 2012
- Modeling of Trabecular Architecture as Result of an Optimal Control Procedure2012
- Walking while resisting a perturbation: Effects on ankle dorsiflexor activation during swing and potential for rehabilitationGait & Posture, 2011
- Intelligently controllable Ankle Foot Orthosis (I-AFO) and its application for a patient of Guillain-Barre syndrome2009
- Prediction of micromotion initiation of an implanted femur under physiological loads and constraints using the finite element methodProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2009
- Surface electrical stimulation for foot drop: Control aspects and walking performanceJournal of Automatic Control, 2008
- A Multicenter Trial of a Footdrop Stimulator Controlled by a Tilt SensorNeurorehabilitation and Neural Repair, 2006
- Modulation in spinal circuits and corticospinal connections following nerve stimulation and operant conditioning2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2006
- BIONic WalkAide for correcting foot dropIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2005
- A reliable gait phase detection systemIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2001