A Biomechanical Study Comparing Helical Blade with Screw Design for Sliding Hip Fixations of Unstable Intertrochanteric Fractures
Open Access
- 1 January 2013
- journal article
- research article
- Published by Wiley in The Scientific World Journal
- Vol. 2013 (1) , 351936
Abstract
Dynamic hip screw (DHS) is a well‐established conventional implant for treating intertrochanteric fracture. However, revision surgery sometimes still occurs due to the cutting out of implants. A helical blade instead of threaded screw (DHS blade) was designed to improve the fixation power of the osteoporotic intertrochanteric fracture. In this study, the biomechanical properties of DHS blade compared to the conventional DHS were evaluated using an unstable AO/OTA 31‐A2 intertrochanteric fracture model. Fifty synthetic proximal femoral bone models with such configuration were fixed with DHS and DHS blade in five different positions: centre‐centre (CC), superior‐centre (SC), inferior‐center (IC), centre‐anterior (CA), and centre‐posterior (CP). All models had undergone mechanical compression test, and the vertical and rotational displacements were recorded. The results showed that DHS blade had less vertical or rotational displacement than the conventional DHS in CC, CA, and IC positions. The greatest vertical and rotational displacements were found at CP position in both groups. Overall speaking, DHS blade was superior in resisting vertical or rotational displacement in comparison to conventional DHS, and the centre‐posterior position had the poorest performance in both groups.Keywords
This publication has 28 references indexed in Scilit:
- Risk factors in cutout of sliding hip screw in intertrochanteric fractures: an evaluation of 937 patientsInternational Orthopaedics, 2009
- Quantification of cancellous bone-compaction due to DHS® Blade insertion and influence upon cut-out resistanceClinical Biomechanics, 2009
- Anterior or posterior obliquity of the lag screw in the lateral view—Does it affect the sliding characteristics on unstable trochanteric fractures?Injury, 2008
- A prospective, randomised study comparing the percutaneous compression plate and the compression hip screw for the treatment of intertrochanteric fractures of the hipThe Journal of Bone and Joint Surgery. British volume, 2007
- Anterior or posterior obliquity of the lag screw in the lateral view—Does it affect the sliding characteristics on unstable trochanteric fractures?Injury, 2007
- Two-hole side-plate DHS in the treatment of intertrochanteric fracture: Results and complicationsInjury, 2005
- Sliding hip screw fixation of trochanteric hip fractures: Outcome of 1024 proceduresInjury, 2005
- Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hipThe Journal of Bone and Joint Surgery. British volume, 1997
- Dynamic hip screws that failInjury, 1991
- Implant failures in patients with proximal fractures of the femur treated with a sliding screw deviceInjury, 1989