In a series of experimental studies, the bone formation around systematically modified titanium implants is analyzed. In the present study, three different surface modifications were prepared and evaluated. Glow-discharge cleaning and oxidizing resulted in a highly stoichiometric TiO2surface, while a glow-discharge treatment in nitrogen gas resulted in implants with essentially a surface of titanium nitride, covered with a very thin titanium oxide. Finally, hydrogen peroxide treatment of implants resulted in an almost stoichiometric TiO2, rich in hydroxyl groups on the surface. Machined commercially pure titanium implants served as controls. Scanning Auger Electron Spectroscopy, Scanning Electron Microscopy, and Atomic Force Microscopy revealed no significant differences in oxide thickness or surface roughness parameters, but differences in the surface chemical composition and apparent topography were observed. After surface preparation, the implants were inserted in cortical bone of rabbits and evaluated after 1, 3, and 6 weeks. Light microscopic evaluation of the tissue response showed that all implants were in contact with bone and had a large proportion of newly formed bone within the threads after 6 weeks. There were no morphological differences between the four groups. Our study shows that a high degree of bone contact and bone formation can be achieved with titanium implants of different surface composition and topography.Keywords
Funding Information
- King Gustaf V 80-year Fund
- Medical Faculty and the Faculty of Odontology, University of Gothenburg
- Swedish National Association against Rheumatism
- Hjalmar Svensson Research Foundation
- Loo and Hans Osterman Fund
- VINNOVA
- Swedish Medical Research Council
- Biomaterial Consortium
- Nobel Biocare AB, Sweden
This publication has 58 references indexed in Scilit:
- TiNOx coatings on roughened titanium and CoCr alloy accelerate early osseointegration of dental implants in minipigsBone, 2013
- Plasma Treatment Maintains Surface Energy of the Implant Surface and Enhances OsseointegrationInternational Journal of Biomaterials, 2013
- Effect of cleaning and sterilization on titanium implant surface properties and cellular responseActa Biomaterialia, 2012
- Elimination of Biological Contaminations from Surfaces by Plasma Discharges: Chemical SputteringChemphyschem, 2010
- Requirement for both micron- and submicron scale structure for synergistic responses of osteoblasts to substrate surface energy and topographyBiomaterials, 2007
- Surface Properties and Biocompatibility of Nitrided Titanium for Abrasion Resistant Implant Materials.Dental Materials Journal, 2002
- Scanning probe metrologyJournal of Vacuum Science & Technology A, 1992
- Oxidation of titanium nitride in room air and in dry O2Journal of Applied Physics, 1991
- A surface spectroscopic study of nitrogen ion‐implanted Ti and Ti–6Al–4V wear against UHMWPESurface and Interface Analysis, 1990
- SOFT X‐RAY REPLICATION OF BIOLOGICAL MATERIAL — X‐RAY MICROSCOPY AND MICROCHEMICAL ANALYSIS OF CELLSAnnals of the New York Academy of Sciences, 1980