Validation of qualitative microbiological test methods
- 21 November 2014
- journal article
- research article
- Published by Wiley in Pharmaceutical Statistics
- Vol. 14 (2) , 120-128
Abstract
This paper considers a statistical model for the detection mechanism of qualitative microbiological test methods with a parameter for the detection proportion (the probability to detect a single organism) and a parameter for the false positive rate. It is demonstrated that the detection proportion and the bacterial density cannot be estimated separately, not even in a multiple dilution experiment. Only the product can be estimated, changing the interpretation of the most probable number estimator. The asymptotic power of the likelihood ratio statistic for comparing an alternative method with the compendial method, is optimal for a single dilution experiment. The bacterial density should either be close to two CFUs per test unit or equal to zero, depending on differences in the model parameters between the two test methods. The proposed strategy for method validation is to use these two dilutions and test for differences in the two model parameters, addressing the validation parameters specificity and accuracy. Robustness of these two parameters might still be required, but all other validation parameters can be omitted. A confidence interval‐based approach for the ratio of the detection proportions for the two methods is recommended, since it is most informative and close to the power of the likelihood ratio test. Copyright © 2014 John Wiley & Sons, Ltd.Keywords
This publication has 15 references indexed in Scilit:
- A comparison of test statistics for the recovery of rapid growth‐based enumeration testsPharmaceutical Statistics, 2013
- Estimation of the limit of detection for quantal response bioassaysPharmaceutical Statistics, 2011
- The Most Probable Limit of Detection (MPL) for rapid microbiological methodsJournal of Microbiological Methods, 2010
- Production of precise microbiology standards using flow cytometry and freeze dryingCytometry Part A, 2004
- Applied Logistic Regression2000
- A Generalized One-Hit Model for Bioassays of Insect Viruses1993
- Multiple‐dose design and bias‐reducing methods for limiting dilution assaysStatistica Neerlandica, 1990
- Estimating the Number of Organisms in Quantal AssaysApplied Microbiology, 1972
- The Large-Sample Distribution of the Likelihood Ratio for Testing Composite HypothesesThe Annals of Mathematical Statistics, 1938
- A Method of Computing the Effectiveness of an InsecticideJournal of Economic Entomology, 1925