Relative Precision of Inhaler Aerodynamic Particle Size Distribution (APSD) Metrics by Full Resolution and Abbreviated Andersen Cascade Impactors (ACIs): Part 1
- 18 May 2010
- journal article
- Published by Springer Nature in AAPS PharmSciTech
- Vol. 11 (2) , 843-851
Abstract
The purpose of this study was to compare relative precision of two different abbreviated impactor measurement (AIM) systems and a traditional multi-stage cascade impactor (CI). The experimental design was chosen to provide separate estimates of variability for each impactor type. Full-resolution CIs are useful for characterizing the aerosol aerodynamic particle size distribution of orally inhaled products during development but are too cumbersome, time-consuming, and resource-intensive for other applications, such as routine quality control (QC). This article presents a proof-of-concept experiment, where two AIM systems configured to provide metrics pertinent to QC (QC-system) and human respiratory tract (HRT-system) were evaluated using a hydrofluoroalkane-albuterol pressurized metered dose inhaler. The Andersen eight-stage CI (ACI) served as the benchmark apparatus. The statistical design allowed estimation of precision with each CI configuration. Apart from one source of systematic error affecting extra-fine particle fraction from the HRT-system, no other bias was detected with either abbreviated system. The observed bias was shown to be caused by particle bounce following the displacement of surfactant by the shear force of the airflow diverging above the collection plate of the second impaction stage. A procedure was subsequently developed that eliminated this source of error, as described in the second article of this series (submitted to AAPS PharmSciTech). Measurements obtained with both abbreviated impactors were very similar in precision to the ACI for all measures of in vitro performance evaluated. Such abbreviated impactors can therefore be substituted for the ACI in certain situations, such as inhaler QC or add-on device testing.Keywords
This publication has 8 references indexed in Scilit:
- Improved Quality Control Metrics for Cascade Impaction Measurements of Orally Inhaled Drug Products (OIPs)AAPS PharmSciTech, 2009
- The Abbreviated Impactor Measurement (AIM) Concept: Part II—Influence of Evaporation of a Volatile Component—Evaluation with a “Droplet-Producing” Pressurized Metered Dose Inhaler (pMDI)-Based Formulation Containing Ethanol as CosolventAAPS PharmSciTech, 2009
- The Abbreviated Impactor Measurement (AIM) Concept: Part 1—Influence of Particle Bounce and Re-Entrainment—Evaluation with a “Dry” Pressurized Metered Dose Inhaler (pMDI)-Based FormulationAAPS PharmSciTech, 2009
- Minimizing Variability of Cascade Impaction Measurements in Inhalers and NebulizersAAPS PharmSciTech, 2008
- Canadian Standards Association Standard CAN/CSA/Z264.1‐02:2002: A New Voluntary Standard for Spacers and Holding Chambers Used with Pressurized Metered‐Dose InhalersCanadian Respiratory Journal, 2004
- Cascade Impactors for the Size Characterization of Aerosols from Medical Inhalers: Their Uses and LimitationsJournal of Aerosol Medicine, 2003
- Considerations for the Development and Practice of Cascade Impaction Testing, Including a Mass Balance Failure Investigation TreeJournal of Aerosol Medicine, 2003
- The Role of Inertial Particle Collectors in Evaluating Pharmaceutical Aerosol Delivery SystemsJournal of Aerosol Medicine, 1998