Evaluation of the basic properties of the BANGkit™ gel dosimeter
Authors:
- T NakashimaTNT Nakashima
- Y WatanabeYWY Watanabe
- T AkimitsuTAT Akimitsu
- K MatsuuraKMK Matsuura
- M KenjoMKM Kenjo
- Y KaneyasuYKY Kaneyasu
- K WadasakiKWK Wadasaki
- Y HirokawaYHY Hirokawa
- K ItoKIK Ito
- 2 April 2007
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (8) , 2301-2311
Abstract
We evaluated the basic properties of a commercially available BANGkit gel dosimeter, which is a normoxic type of BANG gel. This gel-kit has the same composition as the BANG 3 gel, but is fully oxygenated. To exclude oxygen, oxygen scavenging ascorbic acid and copper sulfate as a catalyst are used. The properties that we examined are the effects of the concentrations of copper sulfate and ascorbic acid on the response, the reproducibility, the long-term stability, the temperature effect at irradiation and the dose-rate effect. In our results, the excellent linear fit of the R2-dose response in a dose range for clinical use and its reproducibility were observed. The precision of a linear fit was preserved for about 3 weeks. The temperature at irradiation showed a significant effect on the dose response. Although the dose-rate dependence in the high-dose range was observed, it was negligible for the clinical dose range up to 270 cGy. In conclusion, this gel dosimeter is thought to be utilizable in clinical practice, while we have to pay attention to the temperature during the entire measurement processes, and additionally there is room for improvement in the linearity and the dose-rate dependence in the high-dose range.Keywords
This publication has 18 references indexed in Scilit:
- Basic investigations on the performance of a normoxic polymer gel with tetrakis‐hydroxy‐methyl‐phosphonium chloride as an oxygen scavenger: Reproducibility, accuracy, stability, and dose rate dependenceMedical Physics, 2006
- The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gelsPhysics in Medicine & Biology, 2006
- A study of a normoxic polymer gel dosimeter comprising methacrylic acid, gelatin and tetrakis (hydroxymethyl) phosphonium chloride (MAGAT)Applied Radiation and Isotopes, 2005
- Variable transformation of calibration equations for radiation dosimetryPhysics in Medicine & Biology, 2005
- MAGIC‐type polymer gel for three‐dimensional dosimetry: Intensity‐modulated radiation therapy verificationMedical Physics, 2003
- Energy and dose rate dependence of BANG‐2 polymer‐gel dosimeterMedical Physics, 2001
- Temperature dependence of polymer‐gel dosimeter nuclear magnetic resonance responseMedical Physics, 2001
- Radiation therapy dosimetry using magnetic resonance imaging of polymer gelsMedical Physics, 1996
- Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeterPhysics in Medicine & Biology, 1994
- NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: A new approach to 3D dosimetry by MRIMagnetic Resonance Imaging, 1993