Developing a MLC modifier program to improve fiducial detection for MV/kV imaging during hypofractionated prostate volumetric modulated arc therapy
Open Access
- 22 May 2019
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 20 (6) , 120-124
Abstract
Purpose: To develop an Eclipse plug‐in (MLC_MODIFIER) that automatically modifies control points to expose fiducials obscured by MLC during VMAT, thereby facilitating tracking using periodic MV/kV imaging.Method: Three‐dimensional fiducial tracking was performed during VMAT by pairing short‐arc (3°) MV digital tomosynthesis (DTS) images to triggered kV images. To evaluate MLC_MODIFIER efficacy, two cohorts of patients were considered. For first 12 patients, plans were manually edited to expose one fiducial marker. Next for 15 patients, plans were modified using MLC_MODIFIER script. MLC_MODIFIER evaluated MLC apertures at appropriate angles for marker visibility. Angles subtended by control points were compressed and low‐dose “imaging” control points were inserted and exposed one marker with 1 cm margin. Patient's images were retrospectively reviewed to determine rate of MV registration failures. Failure categories were poor DTS image quality, MLC blockage of fiducials, or unknown reasons. Dosimetric differences in rectum, bladder, and urethra D1 cc, PTV maximum dose, and PTV dose homogeneity (PTV HI) were evaluated. Statistical significance was evaluated using Fisher's exact and Student's t test.Result: Overall MV registration failures, failures due to poor image quality, MLC blockage, and unknown reasons were 33% versus 8.9% (P < 0.0001), 8% versus 6.4% (P < 0.05), 13.6% versus 0.1% (P < 0.0001), and 7.6% versus 2.4% (P < 0.0001) for manually edited and MLC_MODIFIER plans, respectively. PTV maximum and HI increased on average from unmodified plans by 2.1% and 0.3% (P < 0.004) and 22.0% and 3.3% (P < 0.004) for manually edited and MLC_MODIFIED plans, respectively. Changes in bladder, rectum, and urethra D1CC were similar for each method and less than 0.7%.Conclusion: Increasing fiducial visibility via an automated process comprised of angular compression of control points and insertion of additional “imaging” control points is feasible. Degradation of plan quality is minimal. Fiducial detection and registration success rates are significantly improved compared to manually edited apertures.Keywords
This publication has 5 references indexed in Scilit:
- Review of Real-Time 3-Dimensional Image Guided Radiation Therapy on Standard-Equipped Cancer Radiation Therapy Systems: Are We at the Tipping Point for the Era of Real-Time Radiation Therapy?International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2018
- Simultaneous MV‐kV imaging for intrafractional motion management during volumetric‐modulated arc therapy delivery*Journal of Applied Clinical Medical Physics, 2016
- Automatic tracking of arbitrarily shaped implanted markers in kilovoltage projection images: A feasibility studyMedical Physics, 2014
- Short Arc Digital Tomosynthesis as an Image Enhancement Step for Detection and Localization of Implanted MarkersInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Improved Clinical Outcomes With High-Dose Image Guided Radiotherapy Compared With Non-IGRT for the Treatment of Clinically Localized Prostate CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012