Control of FES thumb force using slip information obtained from the cutaneous electroneurogram in quadriplegic man
Authors:
- M. HauglandMHM. Haugland
- A. LickelALA. Lickel
- J. HaaseJHJ. Haase
- T. SinkjaerTST. Sinkjaer
- 1 June 1999
- journal article
- case report
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 7 (2) , 215-227
Abstract
A tetraplegic volunteer was implanted with percutaneous intramuscular electrodes in hand and forearm muscles. Furthermore, a sensory nerve cuff electrode was implanted on the volar digital nerve to the radial side of the index finger branching off the median nerve. In laboratory experiments a stimulation system was used to produce a lateral grasp (key grip) while the neural activity was recorded with the cuff electrode. The nerve signal contained information that could be used to detect the occurrence of slips and further to increase stimulation intensity to the thumb flexor/adductor muscles to stop the slip. Thereby the system provided a grasp that could catch an object if it started to slip due to, for example, decreasing muscle force or changes in load forces tangential to the surface of the object. This method enabled an automatic adjustment of the stimulation intensity to the lowest possible level without loosing the grip and without any prior knowledge about the strength of the muscles and the weight and surface texture of the object.Keywords
This publication has 32 references indexed in Scilit:
- Control with Natural Sensors2000
- Restoration of Lateral Hand Grasp Using Natural Sensors1996
- Closed-loop control of ankle position using muscle afferent feedback with functional neuromuscular stimulationIEEE Transactions on Biomedical Engineering, 1996
- Design and clinical application of a double helix electrode for functional electrical stimulationIEEE Transactions on Biomedical Engineering, 1994
- Sensory nerve recording for closed-loop control to restore motor functionsIEEE Transactions on Biomedical Engineering, 1993
- Development of percutaneous intramuscular electrode for multichannel FES systemIEEE Transactions on Biomedical Engineering, 1989
- Roles of glabrous skin receptors and sensorimotor memory in automatic control of precision grip when lifting rougher or more slippery objectsExperimental Brain Research, 1984
- Tactile sensibility in the human hand: relative and absolute densities of four types of mechanoreceptive units in glabrous skin.The Journal of Physiology, 1979
- Stimulus—response functions of rapidly adapting mechanoreceptors in the human glabrous skin areaThe Journal of Physiology, 1973
- Activity from skin mechanoreceptors recorded percutaneously in awake human subjectsExperimental Neurology, 1968