Multicenter study to compare the diagnostic performance of CLIA vs. FEIA transglutaminase IgA assays for the diagnosis of celiac disease
Open Access
- 5 January 2023
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 61 (8) , 1446-1454
Abstract
Objectives: Celiac disease (CD) is an immune-mediated enteropathy driven by gluten intake. Presence of tTG-IgA antibodies is important for the diagnosis. However, different tTG-IgA assays are used and test performance may vary. Therefore, a retrospective multicenter study was performed to compare the diagnostic performance of three assays. Methods: The fluorescence enzyme-linked immunoassay (FEIA) EliA Celikey IgA (Phadia), the chemiluminescence immunoassays (CLIA) h-tTG IgA QUANTA Flash® (Inova Diagnostics) and the anti-tTG ChLIA IgA (Euroimmun) were compared. Diagnostic samples from CD cases (95 adults; 65 children) and controls (479 adults; 253 children) were included. Samples were blinded and reanalyzed on all platforms. Results: A high quantitative correlation between platforms was found (p<0.0001). Both CLIA were more sensitive (adults 100%; children 100%) compared to the FEIA (adults 88.4%; children 96.6%). Specificity of all assays was high (≥97.6%) with the FEIA having the highest specificity. A cut-off based on receiver operator characteristic analysis (6.5 U/mL) improved the sensitivity of the FEIA (adults 95.8%; children 100%) without affecting specificity. Cut-off values for the CLIA assays did not need further optimization. With the FEIA, 71% of pediatric cases had a tTG-IgA level ≥10× upper limit of normal compared to 91 and 92% with QUANTA Flash and ChLIA, respectively. Conclusions: All platforms have high diagnostic accuracy. The CLIA assays are more sensitive compared to the FEIA assay. A lower cut-off for the FEIA improves diagnostic performance, particularly in adult cases that, as demonstrated in this study, present with lower tTG-IgA levels compared to pediatric cases.Keywords
Funding Information
- JD reports consultancy
- Thermo Fisher Scientific, and Euroimmun
This publication has 30 references indexed in Scilit:
- T Cells in Celiac DiseaseThe Journal of Immunology, 2017
- Clinical utility of quantitative multi-antibody Polycheck immunoassays in the diagnosis of coeliac diseaseWorld Journal of Gastrointestinal Pharmacology and Therapeutics, 2016
- Are ESPGHAN “Biopsy-Sparing” Guidelines for Celiac Disease also Suitable for Asymptomatic Patients?American Journal of Gastroenterology, 2015
- Antibodies in the Diagnosis of Coeliac Disease: A Biopsy-Controlled, International, Multicentre Study of 376 Children with Coeliac Disease and 695 ControlsPLOS ONE, 2014
- Analytical and Clinical Comparison of Two Fully Automated Immunoassay Systems for the Diagnosis of Celiac DiseaseJournal of Immunology Research, 2014
- High-titre circulating tissue transglutaminase-2 antibodies predict small bowel villous atrophy, but decision cut-off limits must be locally validatedClinical and Experimental Immunology, 2013
- ESPGHAN Guidance on Coeliac Disease 2012Journal of Pediatric Gastroenterology and Nutrition, 2012
- Serological diagnosis of celiac disease: Comparative analysis of different strategiesClinica Chimica Acta, 2012
- European Society for Pediatric Gastroenterology, Hepatology, and Nutrition Guidelines for the Diagnosis of Coeliac DiseaseJournal of Pediatric Gastroenterology and Nutrition, 2012
- Diagnostic performance of IgG anti-deamidated gliadin peptide antibody assays is comparable to IgA anti-tTG in celiac diseaseClinica Chimica Acta, 2010