Performance evaluation of the Abbott CELL‐DYN Emerald for use as a bench‐top analyzer in a research setting

Authors:

Abstract
Summary: Introduction: The CELL‐DYN Emerald is a compact bench‐top hematology analyzer that can be used for a three‐part white cell differential analysis. To determine its utility for analysis of human and mouse samples, we evaluated this machine against the larger CELL‐DYN Sapphire and Sysmex XT2000iV hematology analyzers.Methods: 120 human (normal and abnormal) and 30 mouse (normal and abnormal) samples were analyzed on both the CELL‐DYN Emerald and CELL‐DYN Sapphire or Sysmex XT2000iV analyzers. For mouse samples, the CELL‐DYN Emerald analyzer required manual recalibration based on the histogram populations.Results: Analysis of the CELL‐DYN Emerald showed excellent precision, within accepted ranges (white cell count CV% = 2.09%; hemoglobin CV% = 1.68%; platelets CV% = 4.13%). Linearity was excellent (R2 ≥ 0.99), carryover was minimal (<1%), and overall interinstrument agreement was acceptable for both human and mouse samples. Comparison between the CELL‐DYN Emerald and Sapphire analyzers for human samples or Sysmex XT2000iV analyzer for mouse samples showed excellent correlation for all parameters.Conclusion: The CELL‐DYN Emerald was generally comparable to the larger reference analyzer for both human and mouse samples. It would be suitable for use in satellite research laboratories or as a backup system in larger laboratories.