Absorbable magnesium-based stent: physiological factors to consider for in vitro degradation assessments
Open Access
- 6 January 2015
- journal article
- review article
- Published by Oxford University Press (OUP) in Regenerative Biomaterials
- Vol. 2 (1) , 59-69
Abstract
Absorbable metals have been widely tested in various in vitro settings using cells to evaluate their possible suitability as an implant material. However, there exists a gap between in vivo and in vitro test results for absorbable materials. A lot of traditional in vitro assessments for permanent materials are no longer applicable to absorbable metallic implants. A key step is to identify and test the relevant microenvironment and parameters in test systems, which should be adapted according to the specific application. New test methods are necessary to reduce the difference between in vivo and in vitro test results and provide more accurate information to better understand absorbable metallic implants. In this investigative review, we strive to summarize the latest test methods for characterizing absorbable magnesium-based stent for bioabsorption/biodegradation behavior in the mimicking vascular environments. Also, this article comprehensively discusses the direction of test standardization for absorbable stents to paint a more accurate picture of the in vivo condition around implants to determine the most important parameters and their dynamic interactions.Keywords
This publication has 88 references indexed in Scilit:
- Atmospheric corrosion of field-exposed AZ31 magnesium in a tropical marine environmentCorrosion Science, 2013
- DREAMS of a bioabsorbable stent coming trueNature Reviews Cardiology, 2013
- Magnesium alloys: Predicting in vivo corrosion with in vitro immersion testingJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2012
- Biodegradable Metals for Cardiovascular Stent Application: Interests and New OpportunitiesInternational Journal of Molecular Sciences, 2011
- Oxygen, a Key Factor Regulating Cell Behavior during Neurogenesis and Cerebral DiseasesFrontiers in Molecular Neuroscience, 2011
- Stress corrosion cracking of rare-earth containing magnesium alloys ZE41, QE22 and Elektron 21 (EV31A) compared with AZ80Materials Science and Engineering: A, 2008
- Critical assessment of arterial transport modelsInternational Journal of Heat and Mass Transfer, 2008
- Corrosion and corrosion testing of magnesium alloysMaterials and Corrosion, 2007
- Modeling of low-density lipoprotein (LDL) transport in the artery—effects of hypertensionInternational Journal of Heat and Mass Transfer, 2006
- Relation between Cerebral Blood Flow and Metabolism Explained by a Model of Oxygen ExchangeJournal of Cerebral Blood Flow & Metabolism, 2003