Developments in Sample Combustion to Carbon Dioxide, and in the Oxford AMS Carbon Dioxide Ion Source System
Authors:
- 1 January 1992
- journal article
- Published by Cambridge University Press (CUP) in Radiocarbon
- Vol. 34 (3) , 306-311
Abstract
We describe the operation of a commercial system as modified for preparation of CO2 for the CO2 source. AMS samples are automatically combusted in a CHN analyzer, and stable isotope measurements are made on line. We describe the performance of this equipment, with reference to yield, sample contamination, memory effect, accuracy of isotope measurement, convenience and cost. We discuss the current status of dating using the CO2 source. This is the only source in operation at Oxford, and has been in routine dating since September 1989. We assess the practicalities of operation, including the latest measurements on background, memory, sample-size requirements and operating schedules. We also describe modifications to the sputter beam optics and to the gas handling systems.Keywords
This publication has 5 references indexed in Scilit:
- A gaseous ion source for routine AMS radiocarbon datingNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1990
- A CO2 negative ion source for 14C datingNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1989
- Use of the CO2 Source in Radiocarbon Dating by AMSRadiocarbon, 1989
- A gas ion source for radiocarbon datingNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1987
- A versatile high intensity negative ion sourceNuclear Instruments and Methods in Physics Research, 1983