Evaluation of a new fluorescence-based device in the detection of incipient occlusal caries lesions
- 11 May 2012
- journal article
- research article
- Published by Springer Nature in Lasers in Medical Science
- Vol. 28 (1) , 193-201
Abstract
The aim of this in vitro study was to evaluate the performance of various contemporary detection tools at incipient occlusal caries. Thirty eight freshly extracted posterior teeth with non-cavitated occlusal caries were subjected to clinical examination and coding according to ICDAS criteria (n = 38). Standardized fluorescence images were taken with VistaProof (Dürr Dental, Bietigheim-Bissingen, Germany). Furthermore, the teeth were calculated with a laser fluorescence pen device (DIAGNOdent pen, Kavo, Biberach, Germany). In all detection methods, assessment of caries was performed twice (with 1-week interval) by two calibrated examiners. Finally, the caries lesions were validated in tooth sections by histological examination. The intra-examiner and inter-examiner kappa coefficient, sensitivity, and specificity were determined for all detection methods at enamel lesions (D1 threshold). Degrees of agreement of each method with the histological status were calculated using receiver operating characteristic statistics and the area under curve (Az values). The kappa intra-examiner/inter-examiner coefficient values (mean ± SD) were 0.74 ± 0.04/0.73 ± 0.07, 0.87 ± 0.04/0.82 ± 0.07, and 0.91 ± 0.06/0.83 ± 0.08 for clinical examination, DIAGNOdent pen, and VistaProof, respectively. The sensitivity for ICDAS was 0.80–0.86, for DIAGNOdent pen was 0.66–0.75, and for the VistaProof device was 0.97. The specificity for all the detection methods were 0.5 (0.02–0.99). The accuracy value for ICDAS was 0.76–0.81, for DIAGNOdent pen 0.66–0.71, and for the VistaProof device 0.92–0.95. The Az values (mean ± SD) were 0.431 ± 0.187, 0.583 ± 0.215, and 0.486 ± 0.207 for ICDAS, DIAGNOdent pen, and VistaProof examination, respectively. No significant differences in Az values were noted among the methods. All detection methods were presented with high inter-examiner and intra-examiner agreement. The new VistaProof device showed the best sensitivity, while DIAGNOdent pen demonstrated the worst one. Specificities were the same for all detection methods. Moreover, they presented the same performance in detection of incipient occlusal caries.Keywords
This publication has 33 references indexed in Scilit:
- Comparative In Vitro Validation of VistaProof and DIAGNOdent Pen for Occlusal Caries Detection in Permanent TeethOperative Dentistry, 2012
- Influence of Different Professional Prophylactic Methods on Fluorescence Measurements for Detection of Occlusal CariesCaries Research, 2011
- Performance of a fluorescence camera for detection of occlusal caries in vitroOdontology, 2011
- Clinical performance of a new laser fluorescence device for detection of occlusal caries lesions in permanent molarsJournal of Dentistry, 2008
- In vitro comparison of laser fluorescence performance with visual examination for detection of occlusal caries in permanent and primary molarsLasers in Medical Science, 2008
- The International Caries Detection and Assessment System (ICDAS): an integrated system for measuring dental cariesCommunity Dentistry and Oral Epidemiology, 2007
- Clinical evaluation of diagnodent in detection of occlusal caries in children.2006
- Occlusal caries diagnosis in permanent teeth: an in vitro studyBrazilian Oral Research, 2005
- Performance and Reproducibility of a Laser Fluorescence System for Detection of Occlusal Caries in vitroCaries Research, 1999
- Reproducibility and Accuracy of Three Methods for Assessment of Demineralization Depth on the Occlusal Surface: An in vitro ExaminationCaries Research, 1997