Staple Line Failure Using the Proximate® 100 mm Linear Cutter
Authors:
- F. SafiruddinFSF. Safiruddin
- M. I. van Berge HenegouwenM. I. van Berge Henegouwen
- J. F. M. SlorsJSJ. F. M. Slors
- W. A. BemelmanWBW. A. Bemelman
Open Access
- 1 August 2008
- journal article
- case report
- Published by Wolters Kluwer Health in Diseases of the Colon & Rectum
- Vol. 51 (8) , 1275-1278
Abstract
An incomplete linear staple line that was discovered during the stapling of an ileal pouch alerted us to evaluate potential usage concerns with linear cutters. This study was designed to assess the integrity of the staple line of three different sizes of linear staplers. In an animal model three different lengths of linear cutters (Proximate®, Ethicon Endo-Surgery) were used to cross-staple and transect the large bowel of one pig to check for the integrity of the proximal end of the staple line. Cross-stapling and transecting across the pig's large bowel demonstrated that if the tissue is advanced up to the highest number on the scale of the 100 mm stapling device, insufficient overlap between the proximal end of the staple line and the proximal end of the cut line occur. Although a more than 100 mm staple line is delivered, the 100 mm cutter may not produce a double-staggered row of staples at the most proximal end of the staple line if the tissue is advanced past the 9.5 cm mark. Ethicon Endo-Surgery has agreed to add indicator markers to the scale label on the instrument to provide the user with additional guidance for tissue placement.Keywords
This publication has 5 references indexed in Scilit:
- Risk factors for anastomotic leakage after anterior resection of the rectumColorectal Disease, 2004
- The Science of Stapling and LeaksObesity Surgery, 2004
- Surgical stapler-associated fatalities and adverse events reported to the Food and Drug Administration1,2Journal of the American College of Surgeons, 2004
- Predicting risk and diminishing the consequences of anastomotic dehiscence following rectal resectionColorectal Disease, 2003
- Colorectal Surgery and Anastomotic LeakageDigestive Surgery, 2002