Abstract
Background  The surface topographical characterization of bone‐anchored implants has been recommended to be based on amplitude, spatial, and hybrid parameters. There are also functional parameters that have the potential to describe characteristics important for a specific application.Purpose  The aim of the present study was to evaluate if parameters that have been described as functional in engineering applications are also relevant in the topographical characterization of bone‐anchored implants.Materials and Methods  The surface topography of threaded titanium implants with different surface roughness (Sa, Sds, and Sdr) was analyzed with an optical interferometer, and five candidating functional parameters (Sbi, Sci, Svi, Sm, and Sc) were calculated. Examples of the same parameters for five commercially available dental implants were also calculated.Results  The highest core fluid retention index (Sci) was displayed by the turned implants, followed by fixtures blasted with 250‐ and 25‐μm particles, respectively. Fixtures blasted with 75‐μm Al2O3 particles displayed the lowest Sci value. This is the inverse order of the bone biological ranking based on earlier in vivo studies with the experimental surfaces included in the present study.Conclusion  A low core fluid retention index (Sci) seems favorable for bone‐anchored implants. Therefore, it is suggested to include Sci to the set of topographical parameters for bone‐anchored implants to possibly predict the biological outcome.