Correcting frequency and phase offsets in MRS data using robust spectral registration
- 12 July 2020
- journal article
- editorial
- Published by Wiley in NMR in Biomedicine
- Vol. 33 (10) , e4368
Abstract
An algorithm for retrospective correction of frequency and phase offsets in MRS data is presented. The algorithm, termed robust spectral registration (rSR), contains a set of subroutines designed to robustly align individual transients in a given dataset even in cases of significant frequency and phase offsets or unstable lipid contamination and residual water signals. Data acquired by complex multiplexed editing approaches with distinct subspectral profiles are also accurately aligned. Automated removal of unstable lipid contamination and residual water signals is applied first, when needed. Frequency and phase offsets are corrected in the time domain by aligning each transient to a weighted average reference in a statistically optimal order using nonlinear least‐squares optimization. The alignment of subspectra in edited datasets is performed using an approach that specifically targets subtraction artifacts in the frequency domain. Weighted averaging is then used for signal averaging to down‐weight poorer‐quality transients. Algorithm performance was assessed on one simulated and 67 in vivo pediatric GABA‐/GSH‐edited HERMES datasets and compared with the performance of a multistep correction method previously developed for aligning HERMES data. The performance of the novel approach was quantitatively assessed by comparing the estimated frequency/phase offsets against the known values for the simulated dataset or by examining the presence of subtraction artifacts in the in vivo data. Spectral quality was improved following robust alignment, especially in cases of significant spectral distortion. rSR reduced more subtraction artifacts than the multistep method in 64% of the GABA difference spectra and 75% of the GSH difference spectra. rSR overcomes the major challenges of frequency and phase correction.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (P41 EB015909, R01 EB016089, R01 EB023963)
- National Institute of Mental Health (R01 MH106564)
- National Institute of Neurological Disorders and Stroke (R01 NS096207)
This publication has 33 references indexed in Scilit:
- Subtraction artifacts and frequency (Mis‐)alignment in J‐difference GABA editingJournal of Magnetic Resonance Imaging, 2012
- Correction of frequency drifts induced by gradient heating in 1H spectra using interleaved reference spectroscopyJournal of Magnetic Resonance Imaging, 2011
- Single‐voxel MRS with prospective motion correction and retrospective frequency correctionNMR in Biomedicine, 2010
- A practical guide to robust detection of GABA in human brain by J-difference spectroscopy at 3 T using a standard volume coilMagnetic Resonance Imaging, 2007
- A new general-purpose fully automatic baseline-correction procedure for 1D and 2D NMR dataJournal of Magnetic Resonance, 2006
- New method for the simultaneous detection of metabolites and water in localized in vivo 1H nuclear magnetic resonance spectroscopyMagnetic Resonance in Medicine, 2005
- A Perfect SmootherAnalytical Chemistry, 2003
- Improved Baseline Recognition and Modeling of FT NMR SpectraJournal of Magnetic Resonance, 2000
- Wavelet Transform: A Method for Derivative Calculation in Analytical ChemistryAnalytical Chemistry, 1998
- Accurate quantification of in vivo31P NMR signals using the variable projection method and prior knowledgeMagnetic Resonance in Medicine, 1988