Rapid antibiotic susceptibility testing in blood culture diagnostics performed by direct inoculation using the VITEK®-2 and BD Phoenix™ platforms
Authors:
- 24 January 2019
- journal article
- research article
- Published by Springer Nature in European Journal of Clinical Microbiology & Infectious Diseases
- Vol. 38 (3) , 471-478
Abstract
Early availability of microbiological results can improve treatment decisions of patients suffering from bloodstream infections. Direct inoculation of automated susceptibility testing (AST) platforms is an approach to shorten time-to-result in blood culture diagnostics. We performed a comparative evaluation of the two commercial AST systems VITEK®-2 and BD Phoenix™ for the direct inoculation with blood culture samples. Furthermore, two different methods of sample preparation were compared in this study. Positive blood cultures were prepared for direct inoculation by use of serum separator tubes and twofold centrifugation. AST was performed with the VITEK®-2 and the BD Phoenix™ system by the standard method according to the manufacturer’s recommendations using subcultures on solid media and by direct inoculation of blood culture samples. A hundred clinical samples from blood cultures were included in this study. Rapid AST by direct inoculation showed inter-test agreement rates ranging from 92.45 to 97.7%. Comparing both AST platforms, the VITEK®-2 system demonstrated a higher test accuracy for direct inoculation. No relevant difference was observed for the two different sample preparation methods. Direct inoculation is an easy and inexpensive approach to obtain early full panel phenotypic AST results in blood culture diagnostics. Sample preparation is sufficiently performed by a simple centrifugation method. Both commercial platforms, the VITEK®-2 and the BD Phoenix™, have proven suitable for the use of direct inoculation.Keywords
This publication has 27 references indexed in Scilit:
- A simple blood-culture bacterial pellet preparation for faster accurate direct bacterial identification and antibiotic susceptibility testing with the VITEK 2 systemJournal of Medical Microbiology, 2013
- Prospective Evaluation of a Matrix-Assisted Laser Desorption Ionization–Time of Flight Mass Spectrometry System in a Hospital Clinical Microbiology Laboratory for Identification of Bacteria and Yeasts: a Bench-by-Bench Study for Assessing the Impact on Time to Identification and Cost-EffectivenessJournal of Clinical Microbiology, 2012
- Strategy for Rapid Identification and Antibiotic Susceptibility Testing of Gram-Negative Bacteria Directly Recovered from Positive Blood Cultures Using the Bruker MALDI Biotyper and the BD Phoenix SystemJournal of Clinical Microbiology, 2012
- Sepsis, severe sepsis and septic shock: changes in incidence, pathogens and outcomesExpert Review of Anti-infective Therapy, 2012
- Comparative evaluation of Vitek 2 identification and susceptibility testing of Gram-negative rods directly and isolated from BacT/ALERT-positive blood culture bottlesEuropean Journal of Clinical Microbiology & Infectious Diseases, 2011
- Rapid identification and antimicrobial susceptibility testing of Gram-positive cocci in blood cultures by direct inoculation into the BD Phoenix systemClinical Microbiology and Infection, 2010
- Performance of VITEK-2 Compact and overnight MicroScan panels for direct identification and susceptibility testing of Gram-negative bacilli from positive FAN BacT/ALERT blood culture bottlesClinical Microbiology and Infection, 2010
- Rapid identification and antimicrobial susceptibility profiling of Gram-positive cocci in blood cultures with the Vitek 2 systemEuropean Journal of Clinical Microbiology & Infectious Diseases, 2009
- Nosocomial Bloodstream Infections in US Hospitals: Analysis of 24,179 Cases from a Prospective Nationwide Surveillance StudyClinical Infectious Diseases, 2004
- Current Trends in the Epidemiology of Nosocomial Bloodstream Infections in Patients with Hematological Malignancies and Solid Neoplasms in Hospitals in the United StatesClinical Infectious Diseases, 2003