Determination of coenzyme Q10 in human breast milk by high‐performance liquid chromatography

Authors:

Abstract
An isocratic HPLC method was developed for the determination of coenzyme Q10 (CoQ10) in human breast milk. After a single‐step liquid–liquid extraction, the milk extract was injected directly into the HPLC system. The analytical method is based on pre‐column inline treatment of CoQ10. Chromatographic separation of CoQ10 and coenzyme Q9 (CoQ9) internal standard was achieved using a reversed‐phase Microsorb‐MV C18 analytical column. CoQ10 and CoQ9 were monitored by an electrochemical detector (ECD). An excellent linearity (r = 0.999) was observed for CoQ10 in the concentration range 0.06–2.5 µmol L−1 in breast milk. The limit of quantitation (LOQ) was 60 nmol L−1. Coefficients of variations (CVs) for intra‐day and inter‐day assay precisions were less than 5%. A total of 194 breast milk samples were analyzed for the CoQ10 concentration; the mean value was 0.32 ± 0.21 µmol L−1. Copyright © 2006 John Wiley & Sons, Ltd.

This publication has 25 references indexed in Scilit: