A Custom-made Pupillometer System for Characterizing Pupillary Light Response
Open Access
- 4 September 2018
- journal article
- research article
- Published by Galenos Yayinevi in Turkish Journal of Ophthalmology
- Vol. 48 (4) , 185-189
Abstract
Objectives: This paper presents the design and construction of a viable pupillometer system and demonstrates its merits with extensive validation tests. Materials and Methods: A web camera was modified by removing its infrared filter and mounted on a chin rest. Light emitting diodes (LEDs) operating at infrared and visible spectra were integrated to provide background and light stimulus, respectively. The LEDs were controlled by a microprocessor board. Stimulation was presented using a periodic paradigm with variable period and duty cycle. Videos of both pupils were recorded at 30 frames/second and processed offline using software developed in-house. The overall system was validated with data gathered from individuals with healthy vision under different stimulation paradigms. Temporal variations in pupil size were determined and analyzed statistically. Results: The analysis revealed that the pupil sizes were accurately measured from the video frames provided that reflections from both infrared and visible lights remain outside the pupil. The system achieved moderate to excellent repeatability scores (87.8 and 86.8% for short 1 second and long 2 second pulses, respectively), which demonstrated its effectiveness and confirmed that it can be used reliably as a pupillometer. Conclusion: The proposed pupillometer system produces useful, quantitative data characterizing pupillary light response. However, further development and implementation are needed to potentially turn it into a low-cost alternative for other studies involving the autonomic nervous system, cognitive function, drug metabolism, pain response, psychology, fatigue, and sleep disorders.Keywords
This publication has 16 references indexed in Scilit:
- A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability ResearchJournal of Chiropractic Medicine, 2016
- Reliability of standard pupillometry practice in neurocritical care: an observational, double-blinded studyCritical Care, 2016
- Effect of Age and Refractive Error on the Melanopsin Mediated Post-Illumination Pupil Response (PIPR)Scientific Reports, 2015
- Pupillometric findings in children with obstructive sleep apneaSleep Medicine, 2015
- The Post-Illumination Pupil Response (PIPR)Investigative Ophthalmology & Visual Science, 2015
- Test-Retest Reliability of Hemifield, Central-Field, and Full-Field Chromatic Pupillometry for Assessing the Function of Melanopsin-Containing Retinal Ganglion CellsInvestigative Ophthalmology & Visual Science, 2015
- Pupillometry: a non-invasive technique for pain assessment in paediatric patientsArchives of Disease in Childhood, 2014
- The Circadian Response of Intrinsically Photosensitive Retinal Ganglion CellsPLOS ONE, 2011
- Chromatic Pupil ResponsesOphthalmology, 2009
- Autonomic nervous system function in narcolepsyJournal of Sleep Research, 1994